Purpose: The Frozen Elephant Trunk (FET) stent is a hybrid endovascular device that may be implemented in the event of an aneurysm or aortic dissection of the aortic arch or superior descending aorta. A Type 1B endoleak can lead to intrasaccular flow during systole and has been identified as a known failure of the FET stent graft. The purpose was to develop in-vitro modelling techniques to enable the investigation of the known failure.
Methods: A silicone aortic phantom and 3D printed surrogate stent graft were manufactured to investigate the haemodynamics of a Type 1B endoleak. Physiological pulsatile flow dynamics distal to the surrogate stent graft were investigated in-vitro using Particle Image Velocimetry (PIV).
Results: PIV captured recirculation zones and an endoleak distal to the surrogate stent graft. The endoleak was developed at the peak of systole and sustained until the onset of diastole. The endoleak was asymmetric, indicating a potential variation in the phantom artery wall thickness or stent alignment. Recirculation was identified on the posterior dorsal line during late systole.
Conclusions: The identification of the Type 1B endoleak proved that in-vitro modelling can be used to investigate complex compliance changes and wall motions. The recirculation may indicate the potential for long term intimal layer inflammatory issues such as atherosclerosis. These results may aid future remediation techniques or stent design. Further development of the methods used in this experiment may assist with the future testing of stents prior to animal or human trial.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s13239-021-00521-2 | DOI Listing |
Clin Endosc
January 2025
Department of Gastroenterology, Ajou University School of Medicine, Suwon, Korea.
Surg Infect (Larchmt)
January 2025
Division of Thoracic and Cardiovascular Surgery, Department of Surgery, Chang Gung Memorial Hospital, Linkou Medical Center, Chang Gung University, Taoyuan City, Taiwan.
Stent graft infection (SGI) caused by complex is rare. The usage of ascending-to-descending aortic bypass (ADAB) in such situations has not yet been fully discussed. Case report and literature review.
View Article and Find Full Text PDFHeart Rhythm O2
December 2024
Vanderbilt Heart and Vascular Institute, Vanderbilt University Medical Center, Nashville, Tennessee.
EJVES Vasc Forum
November 2024
Vascular Surgery, Department of Medical and Surgical Sciences (DIMEC), University of Bologna, Bologna, Italy.
Objective: Paravisceral aortic lesions present significant challenges for endovascular treatment. This retrospective analysis of consecutively treated patients from April 2017 to June 2021 aimed to analyse the outcome of primary intra-operative embolisation of aortic complicated pseudoaneurysms and gutter channels during parallel graft (PG) repair of paravisceral symptomatic aortic pseudoaneurysms.
Methods: Patients with symptomatic pseudoaneurysms of the paravisceral aorta treated with PGs using chimney or periscope configurations were included.
J Clin Exp Hepatol
November 2024
Department of Radiodiagnosis, All India Institute of Medical Sciences, Bhubaneswar, Odisha, 751019, India.
Suspicion of vascular injury during endoscopic retrograde cholangiopancreatography (ERCP) should be raised in the event of intraprocedural bleeding, persistent hyperbilirubinemia, and sepsis despite biliary stenting. Most inadvertent portal vein (PV) cannulations during ERCP are innocuous, and mere withdrawal of guidewire and catheter suffices. However, unintentional PV stenting, particularly with larger metallic stents, increases the likelihood of significant bleeding.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!