PIV Analysis of Haemodynamics Distal to the Frozen Elephant Trunk Stent Surrogate.

Cardiovasc Eng Technol

Department of Mechanical Engineering, University of Canterbury, Private Bag 4800, Christchurch, 8041, New Zealand.

Published: August 2021

Purpose: The Frozen Elephant Trunk (FET) stent is a hybrid endovascular device that may be implemented in the event of an aneurysm or aortic dissection of the aortic arch or superior descending aorta. A Type 1B endoleak can lead to intrasaccular flow during systole and has been identified as a known failure of the FET stent graft. The purpose was to develop in-vitro modelling techniques to enable the investigation of the known failure.

Methods: A silicone aortic phantom and 3D printed surrogate stent graft were manufactured to investigate the haemodynamics of a Type 1B endoleak. Physiological pulsatile flow dynamics distal to the surrogate stent graft were investigated in-vitro using Particle Image Velocimetry (PIV).

Results: PIV captured recirculation zones and an endoleak distal to the surrogate stent graft. The endoleak was developed at the peak of systole and sustained until the onset of diastole. The endoleak was asymmetric, indicating a potential variation in the phantom artery wall thickness or stent alignment. Recirculation was identified on the posterior dorsal line during late systole.

Conclusions: The identification of the Type 1B endoleak proved that in-vitro modelling can be used to investigate complex compliance changes and wall motions. The recirculation may indicate the potential for long term intimal layer inflammatory issues such as atherosclerosis. These results may aid future remediation techniques or stent design. Further development of the methods used in this experiment may assist with the future testing of stents prior to animal or human trial.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s13239-021-00521-2DOI Listing

Publication Analysis

Top Keywords

stent graft
16
type endoleak
12
surrogate stent
12
frozen elephant
8
elephant trunk
8
stent
8
fet stent
8
in-vitro modelling
8
distal surrogate
8
endoleak
6

Similar Publications

Ascending-to-Descending Aortic Bypass of a Thoracic Stent Graft Infection with a Rare Pathogen.

Surg Infect (Larchmt)

January 2025

Division of Thoracic and Cardiovascular Surgery, Department of Surgery, Chang Gung Memorial Hospital, Linkou Medical Center, Chang Gung University, Taoyuan City, Taiwan.

Stent graft infection (SGI) caused by complex is rare. The usage of ascending-to-descending aortic bypass (ADAB) in such situations has not yet been fully discussed. Case report and literature review.

View Article and Find Full Text PDF

Objective: Paravisceral aortic lesions present significant challenges for endovascular treatment. This retrospective analysis of consecutively treated patients from April 2017 to June 2021 aimed to analyse the outcome of primary intra-operative embolisation of aortic complicated pseudoaneurysms and gutter channels during parallel graft (PG) repair of paravisceral symptomatic aortic pseudoaneurysms.

Methods: Patients with symptomatic pseudoaneurysms of the paravisceral aorta treated with PGs using chimney or periscope configurations were included.

View Article and Find Full Text PDF

Suspicion of vascular injury during endoscopic retrograde cholangiopancreatography (ERCP) should be raised in the event of intraprocedural bleeding, persistent hyperbilirubinemia, and sepsis despite biliary stenting. Most inadvertent portal vein (PV) cannulations during ERCP are innocuous, and mere withdrawal of guidewire and catheter suffices. However, unintentional PV stenting, particularly with larger metallic stents, increases the likelihood of significant bleeding.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!