Thermal Analysis of Heat Transfer from Catheters and Implantable Devices to the Blood Flow.

Micromachines (Basel)

Electrical and Computer Engineering Department, University of Alabama at Birmingham, Alabama, AL 35294, USA.

Published: February 2021

AI Article Synopsis

  • The paper discusses the potential risks of heat generation from biomedical devices like implantable and ablation catheters, highlighting that temperatures above 315 K can be harmful.
  • A heat-transfer model is developed to calculate how heat affects human blood under varying flow velocities, using a Non-Newtonian fluid model and Finite Element methods for analysis.
  • The findings show specific limits for heat flux based on blood flow rates, concluding with an equation that predicts temperature changes relative to heat flux and blood velocity at the site of the device.

Article Abstract

Implantable devices, ultrasound imaging catheters, and ablation catheters (such as renal denervation catheters) are biomedical instruments that generate heat in the body. The generated heat can be harmful if the body temperature exceeds the limit of almost 315 K. This paper presents a heat-transfer model and analysis, to evaluate the temperature rise in human blood due to the power loss of medical catheters and implantable devices. The dynamic of the heat transfer is modeled for the blood vessel, at different blood flow velocities. The physics and governing equations of the heat transfer from the implanted energy source to the blood and temperature rise are expressed by developing a Non-Newtonian Carreau-Yasuda fluid model. We used a Finite Element method to solve the governing equations of the established model, considering the boundary conditions and average blood flow velocities of 0-1.4 m/s for the flow of the blood passing over the implanted power source. The results revealed a maximum allowable heat flux of 7500 and 15,000 W/m for the blood flow velocities of 0 and 1.4 m/s, respectively. The rise of temperature around the implant or tip of the catheter is slower and disappeared gradually with the blood flow, which allows a higher level of heat flux to be generated. The results of this analysis are concluded in the equation/correlation T=310+H3000(1+e-7V), to estimate and predict the temperature changes as a function of heat flux, , and the blood flow velocity, , at the implant/catheter location.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7996209PMC
http://dx.doi.org/10.3390/mi12030230DOI Listing

Publication Analysis

Top Keywords

blood flow
24
heat transfer
12
implantable devices
12
flow velocities
12
heat flux
12
blood
10
heat
8
catheters implantable
8
temperature rise
8
governing equations
8

Similar Publications