Cryo-EM grid preparation is an important bottleneck in protein structure determination, especially for membrane proteins, typically requiring screening of a large number of conditions. We systematically investigated the effects of buffer components, blotting conditions and grid types on the outcome of grid preparation of five different membrane protein samples. Aggregation was the most common type of problem which was addressed by changing detergents, salt concentration or reconstitution of proteins into nanodiscs or amphipols. We show that the optimal concentration of detergent is between 0.05 and 0.4% and that the presence of a low concentration of detergent with a high critical micellar concentration protects the proteins from denaturation at the air-water interface. Furthermore, we discuss the strategies for achieving an adequate ice thickness, particle coverage and orientation distribution on free ice and on support films. Our findings provide a clear roadmap for comprehensive screening of conditions for cryo-EM grid preparation of membrane proteins.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7900225PMC
http://dx.doi.org/10.1016/j.isci.2021.102139DOI Listing

Publication Analysis

Top Keywords

cryo-em grid
12
membrane proteins
12
grid preparation
12
preparation membrane
8
concentration detergent
8
proteins
5
grid optimization
4
membrane
4
optimization membrane
4
proteins cryo-em
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?