Background: Community photoscreening for amblyopia had successfully been adopted by many communities, however many clinics curtailed screening as a result of the COVID-19 pandemic. We modified three conventional devices and tested them for outdoor, drive-by socially distanced photoscreening and refraction.
Methods: External frames that provide luminance control and focus distance were fashioned for plusoptiX S12 (Nuremberg, Germany), Adaptica 2WIN in Kaleidos case (Padova, Italy) and the Rebion blinq (Boston, USA). Children were screened by each device and then Retinomax (Righton, Japan) before AAPOS guideline validation.
Results: Eighty-eight children average age 8±7 years had precise refraction and alignment from which 69% AAPOS 2003 risk factors were determined. The sensitivity/specificity/inconclusive rate for plusoptiX was 85%/96%/16%, for 2WIN 79%/89%/5% and for blinq 43%/74%/8%. Blinq improved to 54%/70% when screening for amblyopia ± strabismus. Bland Altman analysis of spherical equivalent showed plusoptiX and 2WIN with less over-minus than Retinomax and J0 and J45 vectors highly reliable for astigmatism determination.
Conclusion: The infrared photorefractors in modified cases reliably screened amblyopia risk factors and refraction. The birefringent scanner provided drive-by results but less reliably with wire-frame opaque case than without the case in a dimly lit room. Modified drive-by photoscreeners could help reduce amblyopia and provide socially distanced refraction during an extended pandemic.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7914108 | PMC |
http://dx.doi.org/10.2147/OPTH.S300871 | DOI Listing |
Indian J Public Health
July 2022
DNB, Sankara Eye Hospitals, Bengaluru, India.
Objectives: Amblyopia is the most common cause of preventable vision impairment in children. This study tried to evaluate the cost-effectiveness of photoscreeners in identifying refractive errors making children at risk of amblyopia.
Materials And Methods: This was designed as a prospective, open-label, randomized controlled study to compare the cost-effectiveness of photoscreeners (2WIN Adaptica and Plusoptix) versus autorefractor and Mohindra retinoscopy to identify the at-risk amblyopia.
Clin Ophthalmol
February 2022
CONESTOGA EYE, Lancaster, PA, USA.
Background: The American Association for Pediatric Ophthalmology and Strabismus (AAPOS) issued a 2021 update of Uniform Validation Guidelines for instrument-based pediatric vision screeners. With each update it is important for each manufacturer to update the Instrument Referral Criteria (IRC) programed into their devices in order to optimize sensitivity and specificity to detect AAPOS criteria.
Methods: De-identified data comparing photoscreening with simultaneous confirmatory examinations constituted separate cohorts for the development of IRC via receiver operating characteristic (ROC) curves.
Clin Ophthalmol
February 2022
Senior Scientist, Department of Ophthalmology, University of Padua, Padua, Padua Province, Italy.
Background: Novel devices target different facets of amblyopia risk factors (ARFs). Through birefringence, the Rebion blinq assesses binocular foveation. The Adaptica 2WIN is a multiradial infrared photorefractor that also estimates ocular alignment.
View Article and Find Full Text PDFClin Ophthalmol
August 2021
Alaska Blind Child Discovery, Alaska Children's EYE & Strabismus, Anchorage, AK, 99508, USA.
Background: Handheld devices can automatically give an estimate of refraction. The established method for refraction comparison using spherical equivalent (M) and J0, J45 vector transformations by Bland-Altman analysis is too complex for non-eye doctors involved with vision screening and remote vision clinics. Therefore, a simpler comparison technique was developed.
View Article and Find Full Text PDFClin Ophthalmol
February 2021
Alaska Blind Child Discovery, Anchorage, AK, USA.
Background: Community photoscreening for amblyopia had successfully been adopted by many communities, however many clinics curtailed screening as a result of the COVID-19 pandemic. We modified three conventional devices and tested them for outdoor, drive-by socially distanced photoscreening and refraction.
Methods: External frames that provide luminance control and focus distance were fashioned for plusoptiX S12 (Nuremberg, Germany), Adaptica 2WIN in Kaleidos case (Padova, Italy) and the Rebion blinq (Boston, USA).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!