Purpose: To evaluate the mechanical stability and complication rates of titanium (Ti) or zirconia (Zr) abutments restored with cantilevered fixed dental prostheses (cFDPs) when supported by 1 or 2 implants.

Materials And Methods: A total of 32 samples were fabricated. Half of the samples received 1 implant, and the other half received 2 implants (Bone Level Implant, Ø 4.1, 13 mm, Straumann) to simulate the clinical situation of 2 or 3 missing maxillary incisors, respectively. Each group was divided into two subgroups (n = 8). Ti-I and Ti-II groups received Ti abutments (Anatomic Abutment, Straumann) supporting 2- or 3-unit metal cFDPs, respectively, while Zr-I and Zr-II groups received Zr abutments (IPS e.max Anatomic Abutment, Straumann). Following the cementation of cFDPs using resin cement (Multilink Automix, Ivoclar Vivadent), the samples were subjected to thermomechanical fatigue load and were subsequently loaded until fracture in a universal testing machine. Following the static loading test, stereomicroscopic analyses (Carl Zeiss) were done to identify the weakest component of the cFDP, abutment, and implant assembly. Mann-Whitney U test was used to evaluate the effect of the number of supporting implants and abutment material on fracture strength values, and the level of statistical significance was set at 5% (α = .05).

Results: All specimens survived the aging, and no screw loosening or fracture was recorded. The mean fracture strength values were 226 N (± 26.45 ), 551.12 N (± 82.19 ), 601 N (± 41.51 ), and 664.5 N (± 37.59 ) for Zr-I, Zr-II, Ti-I, and Ti-II, respectively. The difference between fracture strength values of Ti and Zr groups was significant in favor of Ti abutments (P < .001). The number of supporting implants showed a significantly positive effect on the fracture strength of Zr abutments.

Conclusion: Zirconia abutments demonstrated lower fracture strength values than titanium abutments independent from the number of supporting implants when used under cFDPs. Two-implant supported cFDPs with zirconia abutments have the potential to withstand physiologic forces applied in the anterior region.

Download full-text PDF

Source
http://dx.doi.org/10.11607/ijp.6700DOI Listing

Publication Analysis

Top Keywords

fracture strength
20
strength values
16
zirconia abutments
12
number supporting
12
supporting implants
12
mechanical stability
8
abutments
8
cantilevered fixed
8
fixed dental
8
dental prostheses
8

Similar Publications

Radiologists and other diagnostic imaging specialists play a pivotal role in the management of osteoporosis, a highly prevalent condition of reduced bone strength and increased fracture risk. Bone mineral density (BMD) measurement with dual-energy X-ray absorptiometry (DXA) is a critical component of identifying individuals at high risk for fracture. Strategies to prevent fractures are consolidated in the Osteoporosis Canada clinical practice guideline which was updated in 2023.

View Article and Find Full Text PDF

Introduction: Osteoporosis is a metabolic skeletal disease characterized by low bone mass and strength, and increased risk for fragility fractures. It is a major health issue in aging populations, due to fracture associated increased disability and mortality. Antiresorptive treatments are first line choices in most of the cases.

View Article and Find Full Text PDF

Taking the titanium alloy wing-body connection joint at the rear beam of a certain type of aircraft as the research object, this study analyzed the failure mechanism and verified the structural safety of the wing-body connection joint under actual flight loads. Firstly, this study verified the validity of the loading system and the measuring system in the test system through the pre-test, and the repeatability of the test was analyzed for error to ensure the accuracy of the experimental data. Then, the test piece was subjected to 400,000 random load tests of flight takeoffs and landings, 100,000 Class A load tests, and ground-air-ground load tests, and the test piece fractured under the ground-air-ground load tests.

View Article and Find Full Text PDF

Effect of Chemical Treatment on the Mechanical and Hygroscopic Properties of an Innovative Clay-Sand Composite Reinforced with Fibers.

Materials (Basel)

January 2025

Laboratoire d'Energétique et des Transferts Thermique et Massique (LETTM), Faculté des Sciences de Tunis, Université de Tunis El Manar, Campus Universitaire El-Manar, El Manar, Tunis 2092, Tunisia.

The viability of using fibers as reinforcement material for developing lightweight sustainable non-structural construction materials in compliance with the valorization of local by-products has been investigated in this work. This study aims to investigate the effect of the chemical treatment of fibers on the mechanical and hygric properties of bio-sourced clay-sand- fiber composite. This lightweight specimen has been produced from a mixture of 60% natural clay and 40% sand by mass, as a matrix, and reinforced with different amounts of Juncus fibers.

View Article and Find Full Text PDF

Pultruded carbon fiber-reinforced composites are attractive to the wind energy industry due to the rapid production of highly aligned unidirectional composites with enhanced fiber volume fractions and increased specific strength and stiffness. However, high volume carbon fiber manufacturing remains cost-prohibitive. This study investigates the feasibility of a pultruded low-cost textile carbon fiber-reinforced epoxy composite as a promising material in spar cap production was undertaken based on mechanical response to four-point flexure loading.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!