[Final arrest of a pacemaker after use of electrocautery].

Ann Fr Anesth Reanim

Département d'Anesthésie-Réanimation, CHU, Angers.

Published: June 1988

A case is reported of irreversible damage being caused to a permanent programmable pacemaker by electrocautery used in the epigastric region. The pacemaker was rapidly replaced, and the patient had no adverse effects of this accident. The use of monopolar electrocautery in patients who have one of the new generation of programmable pacemakers is very dangerous. Bipolar forceps can reduce the level of interference between electrocautery units and pacemaker electrodes. With programmable pacemakers, the generator instruction manual should be consulted before surgery, as placing a magnet on the generator may not necessarily convert it to the asynchronous mode. When the use of electrocautery is unavoidable, external cardiac pacing electrodes should be placed on the patient, with an external cardiac pacemaker ready.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0750-7658(88)80147-3DOI Listing

Publication Analysis

Top Keywords

programmable pacemakers
8
external cardiac
8
pacemaker
5
[final arrest
4
arrest pacemaker
4
pacemaker electrocautery]
4
electrocautery] case
4
case reported
4
reported irreversible
4
irreversible damage
4

Similar Publications

Aims: Evidence regarding the safety of early discharge following transcatheter aortic valve implantation (TAVI) is limited. The aim of this study was to evaluate the safety of very early (<24) and early discharge (24-48 h) as compared to standard discharge (>48 h), supported by the implementation of a voice-based virtual assistant using artificial intelligence (AI) and natural language processing.

Methods And Results: Single-arm prospective observational study that included consecutive patients who underwent TAVI in a tertiary hospital in 2023 and were discharged under an AI follow-up programme.

View Article and Find Full Text PDF

Objective: Anorexia nervosa (AN) is an eating disorder characterized by severe weight loss and associated with hyperactivity and circadian rhythm disruption. However, the cellular basis of circadian rhythm disruption is poorly understood. Glial cells in the suprachiasmatic nucleus (SCN), the principal circadian pacemaker, are involved in regulating circadian rhythms.

View Article and Find Full Text PDF

Introduction: Haemodynamic atrioventricular delay (AVD) optimisation has primarily focussed on signals that are not easy to acquire from a pacing system itself, such as invasive left ventricular catheterisation or arterial blood pressure (ABP). In this study, standard clinical central venous pressure (CVP) signals are tested as a potential alternative.

Methods: Sixteen patients with a temporary pacemaker after cardiac surgery were studied.

View Article and Find Full Text PDF

Background: Permanent Pacemaker (PPM) implantation is essential in treating cardiac arrhythmias and conduction disorders, especially in patients with heart failure. Although PPM has been proven to improve quality of life and prolong life expectancy in patients with cardiac conduction disorders, post implantation complications still often occur.

Purpose: This study aimed to identify types of complications and associated predictors in patients undergoing PPM implantation.

View Article and Find Full Text PDF

Nutrient acquisition is crucial for sustaining life. Plants develop beneficial intracellular partnerships with arbuscular mycorrhiza (AM) and nitrogen-fixing bacteria to surmount the scarcity of soil nutrients and tap into atmospheric dinitrogen, respectively. Initiation of these root endosymbioses requires symbiont-induced oscillations in nuclear calcium (Ca) concentrations in root cells.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!