Comaneci Device for Temporary Coiling Assistance for Treatment of Wide-Necked Aneurysms: Initial Case Series and Systematic Literature Review.

World Neurosurg

Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo, New York, USA; Department of Radiology, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo, New York, USA; Canon Stroke and Vascular Research Center, University at Buffalo, Buffalo, New York, USA; Jacobs Institute, Buffalo, New York, USA; Department of Neurosurgery, Gates Vascular Institute at Kaleida Health, Buffalo, New York, USA. Electronic address:

Published: May 2021

Background: Coiling of wide-necked aneurysms requires high-density packing of coils within the aneurysm, which necessitates adequate microcatheter access and navigability. The Comaneci device, introduced in the United States in 2019, is a retrievable stent that can be used as an adjunct to coiling of a wide-necked aneurysm without limiting flow. We present a case series and systematic review of use of this device.

Methods: All cases involving use of the device at our institute between May 1, 2019, and April 30, 2020, were reviewed. A comprehensive systematic review of the literature was conducted using PubMed and EMBASE and a review of eligible article bibliographies.

Results: Five patients underwent Comaneci-assisted wide-necked aneurysm coiling during the study period; 4 were treated via a radial artery approach, and 1 was treated via a femoral artery approach. Two patients presented with subarachnoid hemorrhage: 1 had a ruptured posterior inferior cerebellar artery aneurysm (Hunt-Hess 5, Fisher 4), and 1 had a ruptured middle cerebral artery aneurysm (Hunt-Hess 2, Fisher 1). Mean aneurysmal neck size was 4.4 ± 0.8mm; mean aspect ratio was 1.2 ± 0.3. Raymond-Roy 1 occlusion was achieved in all aneurysms except the posterior inferior cerebellar artery aneurysm. Systematic literature review identified 4 articles that found use of the Comaneci device to treat wide-necked aneurysms to be effective.

Conclusions: This device can be used with transfemoral and transradial approaches, allowing for continued flow through the parent vessel during the coiling procedure while providing a scaffold for dense coiling of the aneurysm and its neck.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2021.02.080DOI Listing

Publication Analysis

Top Keywords

comaneci device
12
wide-necked aneurysms
12
artery aneurysm
12
case series
8
series systematic
8
systematic literature
8
literature review
8
coiling wide-necked
8
wide-necked aneurysm
8
systematic review
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!