Objectives: This study sought to evaluate the effectiveness of His-optimized cardiac resynchronization therapy (HOT-CRT) for reducing left ventricular activation time (LVAT) compared to His bundle pacing (HBP) and biventricular (BiV) pacing (including multipoint pacing [MPP]), using electrocardiographic (ECG) imaging.
Background: HBP may correct bundle branch block (BBB) and has shown encouraging results for providing CRT. However, HBP does not correct BBB in all patients and may be combined with univentricular or BiV fusion pacing to deliver HOT-CRT to maximize resynchronization.
Methods: Nineteen patients with a standard indication for CRT, implanted with HBP without correction of BBB and BiV (n = 14) or right ventricular (n = 5) leads, were prospectively enrolled. Patients underwent ECG imaging while pacing in different configurations using different LV electrodes and at different HBP ventricular pacing (VP) delays. The primary endpoint was reduction in LVAT with HOT-CRT, and the secondary endpoints included various other dys-synchrony measurements including right ventricular activation time (RVAT).
Results: Compared to HBP, HOT-CRT reduced LVAT by 21% (-17 ms [95% confidence interval [CI]: -25 to -9 ms]; p < 0.001) and outperformed BiV by 24% (-22 ms [95% CI: -33 to -10 ms]; p = 0.002) and MPP by 13% (-11 ms [95% CI: -21 to -1 ms]; p = 0.035). Relative to HBP, HOT-CRT also reduced RVAT by 7% (-5 ms [95% CI: -9 to -1 ms; p = 0.035) in patients with right BBB, whereas RVAT was increased by BiV. The other electrical dyssynchrony measurements also improved with HOT-CRT.
Conclusions: HOT-CRT acutely improves ventricular electrical synchrony beyond BiV and MPP. The impact of this finding needs to be evaluated further in studies with clinical follow-up. (Electrical Resynchronization and Acute Hemodynamic Effects of Direct His Bundle Pacing Compared to Biventricular Pacing; NCT03452462).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacep.2020.11.029 | DOI Listing |
J Clin Med
July 2024
Ordensklinikum Linz Elisabethinen, Fadingerstraße 1, 4020 Linz, Austria.
The current gold standard in device therapy for advanced heart failure (HF), which has been firmly established in HF management for more than 25 years, is classical biventricular pacing (BiV-CRT). In the last decade, a new pacing modality called conduction system pacing (CSP) has emerged as a variant for advanced cardiac device therapy. It provides pacing with preserved intrinsic cardiac activation by direct stimulation of the specific cardiac conduction system.
View Article and Find Full Text PDFEur Heart J Suppl
November 2023
National Heart and Lung Institute, Imperial College London, Hammersmith Hospital, Du Cane Road, London W12 0HS, UK.
Patients with heart failure who have a prolonged PR interval are at a greater risk of adverse clinical outcomes than those with a normal PR interval. Potential mechanisms of harm relating to prolonged PR intervals include reduced ventricular filling and also the potential progression to a higher degree heart block. There has, however, been relatively little work specifically focusing on isolated PR prolongation as a therapeutic target.
View Article and Find Full Text PDFJ Cardiovasc Electrophysiol
April 2023
Department of Biomedical Engineering, King's College London, London, UK.
Introduction: Conduction system pacing (CSP), in the form of His bundle pacing (HBP) or left bundle branch pacing (LBBP), is emerging as a valuable cardiac resynchronization therapy (CRT) delivery method. However, patient selection and therapy personalization for CSP delivery remain poorly characterized. We aim to compare pacing-induced electrical synchrony during CRT, HBP, LBBP, HBP with left ventricular (LV) epicardial lead (His-optimized CRT [HOT-CRT]), and LBBP with LV epicardial lead (LBBP-optimized CRT [LOT-CRT]) in patients with different conduction disease presentations using computational modeling.
View Article and Find Full Text PDFEur J Heart Fail
February 2023
National Heart and Lung Institute, Imperial College London, London, UK.
Card Electrophysiol Clin
June 2022
Department of Cardiology, University Hospital of Geneva, Geneva, Switzerland. Electronic address:
Fusion pacing, which exploits conduction via the intrinsic His-Purkinje system, forms the basis of recent cardiac resynchronization therapy (CRT) optimization algorithms. However, settings need to be constantly adjusted to accommodate for changes in AV conduction, and the algorithms are not always available (eg, depending on the device, in case of AV block or with atrial fibrillation). His-optimized cardiac resynchronization therapy (HOT-CRT), and left-bundle branch optimized cardiac resynchronization therapy (LOT-CRT) which combines conduction system pacing with ventricular fusion pacing, provide constant fusion with ventricular activation (irrespective of intrinsic AV conduction).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!