Background: Double J ureteral stents are widely used on urological patients to provide drainage of the upper urinary tract. Unfourtunately, ureteral stents are not free from complications, as bacterial colonization and require a second procedure for removal. The purpose of the current comparative experimental study is to evaluate a new heparin-coated biodegradable antireflux ureteral stent (BraidStent®-H) to prevent urinary bacterial colonization.

Methods: A total of 24 female pigs were underwent determination of bacteriuria and nephrosonographic, endoscopic and contrast fluoroscopy assessment of the urinary tract. Afterward, were randomly assigned animals to Group-I, in which a 5Fr double-pigtail ureteral stent was placed for 6 weeks, or Group-II, in which a BraidStent®-H was placed. Follow-up assessments were performed at 1, 3, 6, 8, 12 weeks. The final follow-up includes the above methods and an exhaustive pathological study of the urinary tract was accomplished after 20 weeks.

Results: Bacteriuria findings in the first 48 h were significant between groups at 6 h and 12 h. Asymptomatic bacteriuria does not reach 100% of the animals in Group-II until 48 h versus Group-I where it appears at 6 h. The weekly bacteriuria mean rate was 27.7% and 44.4% in Group I and II respectively, without statistical significance. In Group II there were no animals with vesicoureteral reflux, with statistical significance at 3 and 6 weeks with Group-I. The 91.2% of stents in Group-II were degraded between 3 and 6 weeks, without obstructive fragments. Distal ureteral peristalsis was maintained in 66.6-75% in Group-II at 1-6 weeks.

Conclusions: The heparin coating of BraidStent® allows an early decrease of bacterial colonization, but its effectiveness is low at the long term. Heparin coating did not affect scheduled degradation rate or size of stents fragments. BraidStent®-H avoids the side effects associated with current ureteral stents, thus should cause less discomfort to patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7916282PMC
http://dx.doi.org/10.1186/s12894-021-00802-xDOI Listing

Publication Analysis

Top Keywords

ureteral stent
12
ureteral stents
12
urinary tract
12
bacterial colonization
8
statistical significance
8
heparin coating
8
ureteral
7
stents
5
comparative assessment
4
assessment biodegradable-antireflux
4

Similar Publications

Purpose: To evaluate stone free rate (SFR) predictivity of three different scoring systems in patients with kidney stones larger than 20 millimeters undergoing retrograde intrarenal surgery(RİRS).

Methods: Digital records of a total of 166 patients were reviewed retrospectively. Epidemiological characteristics (age, gender, medical history) of the patients, stone and affected kidney characteristics (size, volume, location, density, opaque, presence of urinary system anomaly, presence of stones in different calyx, number of stones, lower pole stone, renal infundibulopelvic angle (IPA), renal infundibulopelvic length (RIL), hydronephrosis), and operative characteristics (preoperative ureteral stent, operation duration, postoperative residual fragments, hospitalization time and complications were recorded.

View Article and Find Full Text PDF

Arterio-ureteral fistula (AUF) is a rare condition affecting nearby adult-only patients. Patients usually present with hematuria, often starting as intermittent hematuria which frequently increases, and may lead to hemorrhagic shock. Without rapid diagnosis and prompt treatment, AUF can be lethal.

View Article and Find Full Text PDF

Objective: Retroperitoneal fibrosis (RPF) is a rare disease characterized by the presence of fibroinflammatory tissue that surrounds the abdominal aorta and the iliac arteries and often entraps the ureters. Hydronephrosis is a common complication of RPF, however, its clinical features and outcomes have not been well elucidated.

Methods: A total of 115 RPF-related hydronephrosis patients have been recruited from 9 clinical centers in China since March 2010.

View Article and Find Full Text PDF

Objectives: To present our initial experience of robotic ureteroplasty with lower-lip mucosal graft (LLMG) for treating ureteral stenosis longer than 2 cm and evaluate its feasibility and efficacy.

Materials And Methods: A total of thirteen patients with ureteral stenosis who underwent robotic ureteroplasty with LLMG were retrospectively analyzed. After identification and dissection of the ureteral stenosis segment, the segment was incised longitudinally.

View Article and Find Full Text PDF

Background: This study aimed to investigate initial experiences and outcomes of a retrograde inserted new silicone-covered metallic ureteric stent (Urexel™) for malignant ureteric obstruction.

Methods: We retrospectively reviewed the medical records for 12 consecutive patients who underwent Urexel™ stent placement for malignant ureteric obstruction from March 2020 to March 2021. The Urexel™ stent is a segmental metallic ureteric stent composed of a nitinol mesh covered with a silicone membrane.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!