Tibial Component Migration After Total Knee Arthroplasty With High-Viscosity Bone Cement.

J Arthroplasty

Division of Orthopaedics, Department of Surgery, Dalhousie University and QEII Health Sciences Centre, Nova Scotia Health Authority, Halifax, Nova Scotia, Canada; School of Biomedical Engineering, Faculty of Engineering & Faculty of Medicine, Dalhousie University, Halifax, Nova Scotia, Canada.

Published: June 2021

Background: High-viscosity (HV) bone cements have been formulated to offer potentially advantageous handling characteristics. However, alteration in the handling characteristics could influence implant fixation and survival. The primary objective of this study was to use radiostereometric analysis after total knee arthroplasty to assess the migration of the Triathlon tibial component fixed with HV cement (Simplex HV).

Methods: Twenty-three patients were followed for two years with radiostereometric analysis examinations at 6 visits. Migration was compared with published thresholds and with a control group from a previously published study from the same center using the same implants fixed with a medium viscosity cement. Inducible displacement was assessed, and Oxford 12 Knee Scores and satisfaction were recorded.

Results: Mean maximum total point motion migration reaching 0.40 mm (SD 0.16) at one year, and 0.41mm (SD 0.17) at two years, demonstrating a pattern of stable fixation, below published thresholds of acceptable migration, and not significantly different from the control group. One implant had continuous migration between 1 and 2 years but was clinically asymptomatic. Mean maximum total point motion inducible displacement measured at least one year postoperatively was 0.3 mm (SD 0.12). Mean Oxford 12 Knee Scores improved from 19 (SD 7) preoperatively to 42 (SD 8) 2 years postoperatively.

Conclusions: The use of HV cement demonstrated an acceptable pattern of migration at 2 years, indicating low risk for aseptic loosening.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2021.01.081DOI Listing

Publication Analysis

Top Keywords

tibial component
8
total knee
8
knee arthroplasty
8
high-viscosity bone
8
handling characteristics
8
radiostereometric analysis
8
published thresholds
8
control group
8
inducible displacement
8
oxford knee
8

Similar Publications

Purpose: Double-level osteotomies (DLOs) have shown promising results for knee joint preservation, however, most ultimately progress in terms of degenerative disease resulting in conversion to total knee arthroplasty (TKA). Therefore, the purpose of this study was to examine the time to TKA conversion, long-term clinical outcomes and revision rates of patients who have undergone TKA after prior ipsilateral DLO.

Methods: Patients who underwent simultaneous or staged DLO and subsequently underwent conversion to TKA at a single academic institution from 1997 to 2022 were evaluated.

View Article and Find Full Text PDF

Background And Purpose:  This study updates 2 parallel systematic reviews and meta-analyses from 2012, which established the 1-year radiostereometric (RSA) migration thresholds for tibial components of total knee replacements (TKR) based on the risk of late revision for aseptic loosening from survival studies. The primary aim of this study was to determine the (mis)categorization rate of the 2012 thresholds using the updated review as a validation dataset. Secondary aims were evaluation of 6-month migration, mean continuous (1- to 2-year) migration, and fixation-specific thresholds for tibial component migration.

View Article and Find Full Text PDF

Intraoperatively identified patellofemoral osteoarthritis: no significant impact on short-term outcomes of medial unicompartmental knee arthroplasty.

BMC Musculoskelet Disord

January 2025

Department of Orthopaedic Surgery, The First Affiliated Hospital, Fujian Medical University, No.20, Chazhong road, Fuzhou, 350005, China.

Background And Objective: The efficacy of medial unicompartmental knee arthroplasty (UKA) in patients with intraoperatively identified patellofemoral osteoarthritis (PFOA) has been a subject of debate. This retrospective study aimed to investigate the early outcomes of UKA in patients with varying intraoperative PFOA conditions and to explore the relationship between the location of PFOA and the position of the prosthesis post-UKA. Our aim was to determine whether the presence of PFOA affects the short-term success of medial UKA.

View Article and Find Full Text PDF

The Quantum total ankle prosthesis is a newer Food and Drug Administration (FDA)-approved fourth-generation 2-component, fixed-bearing implant in its first-generation design. The purpose of this study was to evaluate early outcome data and present our initial experience with the Quantum implant with a minimum of a 1-year follow-up. A retrospective, single-centered chart and radiographic review was performed on all patients who underwent total ankle arthroplasty with the Quantum implant from December 2021 to August 2023.

View Article and Find Full Text PDF

Comparison of Biocompatibility of 3D-Printed Ceramic and Titanium in Micropig Ankle Hemiarthroplasty.

Biomedicines

November 2024

Division in Anatomy and Developmental Biology, Department of Oral Biology, Human Identification Research Institute, BK21 FOUR Project, Yonsei University College of Dentistry, 50-1 Yonsei-ro, Seodaemun-gu, Seoul 03722, Republic of Korea.

Background: Ankle arthritis is a common degenerative disease that progresses as cartilage damage in the lower tibia and upper talus progresses, resulting in loss of joint function. In addition to typical arthritis, there is also structural bone loss in the talus due to diseases such as talar avascular necrosis. Total talus replacement surgery is the procedure of choice in end-stage ankle arthritis and consists of a tibial, talar component and an insert.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!