Head-to-head comparison of a Si-photomultiplier-based and a conventional photomultiplier-based PET-CT system.

EJNMMI Phys

Radiation Physics, Department of Hematology, Oncology and Radiation Physics, Skåne University Hospital, 221 85, Lund, Sweden.

Published: February 2021

Background: A novel generation of PET scanners based on silicon (Si)-photomultiplier (PM) technology has recently been introduced. Concurrently, there has been development of new reconstruction methods aimed at increasing the detectability of small lesions without increasing image noise. The combination of new detector technologies and new reconstruction algorithms has been found to increase image quality. However, it is unknown to what extent the demonstrated improvement of image quality is due to scanner hardware development or improved reconstruction algorithms. To isolate the contribution of the hardware, this study aimed to compare the ability to detect small hotspots in phantoms using the latest generation SiPM-based PET/CT scanner (GE Discovery MI) relative to conventional PM-based PET/CT scanner (GE Discovery 690), using identical reconstruction protocols.

Materials And Methods: Two different phantoms (NEMA body and Jasczcak) with fillable spheres (31 μl to 26.5 ml) and varying sphere-to-background-ratios (SBR) were scanned in one bed position for 15-600 s on both scanners. The data were reconstructed using identical reconstruction parameters on both scanners. The recovery-coefficient (RC), noise level, contrast (sphere/background-value), and detectability of each sphere were calculated and compared between the scanners at each acquisition time.

Results: The RC-curves for the NEMA phantom were near-identical for both scanners at SBR 10:1. For smaller spheres in the Jaszczak phantom, the contrast was 1.22 higher for the DMI scanner at SBR 15:1. The ratio decreased for lower SBR, with a ratio of 1.03 at SBR 3.85:1. Regarding the detectability of spheres, the sensitivity was 98% and 88% for the DMI and D690, respectively, for SBR 15:1. For SBR 7.5, the sensitivity was 75% and 83% for the DMI and D690, respectively. For SBR 3.85:1, the sensitivity was 43% and 30% for the DMI and D690, respectively.

Conclusion: Marginally higher contrast in small spheres was seen for the SiPM-based scanner but there was no significant difference in detectability between the scanners. It was difficult to detect differences between the scanners, suggesting that the SiPM-based detectors are not the primary reason for improved image quality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7907292PMC
http://dx.doi.org/10.1186/s40658-021-00366-7DOI Listing

Publication Analysis

Top Keywords

image quality
12
dmi d690
12
reconstruction algorithms
8
pet/ct scanner
8
scanner discovery
8
identical reconstruction
8
sbr
8
sbr 151
8
sbr 3851
8
d690 sbr
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!