A microfluidic cartridge for fast and accurate diagnosis of infections on standard laboratory equipment.

Lab Chip

Hahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany. and Laboratory for MEMS Applications, IMTEK - Department of Microsystems Engineering, University of Freiburg, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.

Published: April 2021

We present a novel centrifugal microfluidic approach for fast and accurate tuberculosis (TB) diagnosis based on the use of standard laboratory equipment. The herein presented workflow can directly be integrated into laboratories with standard equipment and automates complex sample preparation. The system consists of a microfluidic cartridge, a laboratory centrifuge and a standard PCR cycler. The cartridge includes all required reagents and automates collection of bacteria on filter membranes, bacterial lysis, nucleic acid extraction and aliquoting of the DNA extract for PCR analysis. We show that storage of the reagents in aluminium-coated pouches is stable during accelerated storage and transport tests. When the limit of detection was assessed, we found that the cartridge-automated workflow consistently detected 10 CFU ml-1 of mycobacteria in spiked sputum samples. First tests with clinical samples showed a 100% specificity for non-TB specimens. In addition, Mycobacterium tuberculosis (MTB) was re-found in pre-characterized smear microscopy and culture positive sputum samples suggesting a high diagnostic sensitvity. In summary, the novel cartridge-automated workflow enables a flexible and sensitive TB diagnosis without the need to invest in specialized instrumentation.

Download full-text PDF

Source
http://dx.doi.org/10.1039/d1lc00035gDOI Listing

Publication Analysis

Top Keywords

microfluidic cartridge
8
fast accurate
8
standard laboratory
8
laboratory equipment
8
cartridge-automated workflow
8
sputum samples
8
cartridge fast
4
accurate diagnosis
4
diagnosis infections
4
standard
4

Similar Publications

Rapid and accurate molecular diagnostics are crucial for preventing the global spread of emerging infectious diseases. However, the current gold standard for nucleic acid detection, reverse transcription polymerase chain reaction (RT-PCR), relies heavily on traditional magnetic beads or silica membranes for nucleic acid extraction, resulting in several limitations, including time-consuming processes, the need for trained personnel, and complex equipment. Therefore, there is an urgent need for fully integrated nucleic acid detection technologies that are simple to operate, rapid, and highly sensitive to meet unmet clinical needs.

View Article and Find Full Text PDF

Integrating advanced Microfluidic lateral flow systems with a finger-prick blood collection cartridge to create an all-in-one platform for point-of-care diagnostics.

Biosens Bioelectron

March 2025

Laboratory of Advanced Biotechnologies for Health Assessments (Lab-HA), Lassonde School of Engineering, York University, Toronto, ON, M3J 1P3 Canada; Department of Electrical Engineering and Computer Science (EECS), Lassonde School of Engineering, York University, Toronto, ON, M3J 1P3, Canada. Electronic address:

Rapid, point-of-care tests are critical for early diagnosis of disease and detection of biological threats. Lateral flow immunoassays (LFIAs) are well-suited for point-of-care testing due to their ease of use and straightforward readout. However, limitations in sensitivity, quantification, and integration into sample-to-result systems indicate the need for further advancements.

View Article and Find Full Text PDF

Accurate detection of pathogenic nucleic acids is crucial for early diagnosis, effective treatment, and containment of infectious diseases. It facilitates the timely identification of pathogens, aids in monitoring disease outbreaks, and helps prevent the spread of infections within healthcare settings and communities. We developed a multi-layered, paper-based microfluidic and miniaturized electrophoresis system for rapid nucleic acid extraction, separation, amplification, and detection, designed for resource-limited settings.

View Article and Find Full Text PDF

A novel nucleic acid extraction system integrating a switching valve-assisted microfluidic cartridge and a constant pressure pump.

Anal Chim Acta

November 2024

Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, 230031, Anhui, China. Electronic address:

Background: Nucleic acid extraction (NAE) is essential in molecular diagnostics, genetic engineering, and DNA sequencing. While advances in switching valve-assisted, self-contained microfluidic (SM) cartridges have improved the NAE processes, challenges persist in streamlining workflows, reducing processing times, and enhancing multi-target detection. Achieving these objectives requires precise control of on-chip fluid dynamics and efficient transfer of nucleic acid solutions (NAS) to detection areas.

View Article and Find Full Text PDF

The radiometal gallium-68 (Ga-68) has garnered significant interest due to its convenient production via compact and widely available generators and the high performance of Ga-labeled compounds for positron-emission tomography (PET) imaging for cancer diagnosis and management of patients undergoing targeted radionuclide therapy. Given the short half life of Ga-68 (68 min), microfluidic-based radiosynthesis is a promising avenue to establish very rapid, efficient, and routine radiolabeling with Ga-68; however, the typical elution volume of Ga-68 from a generator (4-10 mL) is incompatible with the microliter reaction volumes of microfluidic devices. To bridge this gap, we developed a microscale cartridge-based approach to concentrate Ga-68.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!