Introduction: Two important aspects of the dental operating microscope (DOM) that factor into its overall effectiveness are resolution and depth of field. Therefore, the objective of this study was to evaluate and compare the resolution and depth of field of DOMs from three well-known manufacturers using standardized test targets.

Materials And Methods: A resolution test, using the 1951 USAF Hi-Resolution Target (Edmund Optics, Barrington, NJ), and a depth of field test, using the Depth of Field Target 5-15 (Edmund Optics, Barrington, NJ), were performed by two calibrated observers. Three DOM systems such as Seiler IQ (Seiler Instrument Inc., St. Louis, USA), Global G-Series 6 step (Global Surgical Corp., St. Louis, USA), and Zeiss Extaro 300 (Carl Zeiss Meditec AG, Oberkochen, Germany) were used to compare the resolution and depth of field.

Results: The Zeiss Extaro 300 showed the highest maximum resolution and maximum DOF (64 lp/mm and 17mm, respectively). The Seiler IQ showed the lowest maximum resolution and maximum DOF (35.9 lp/mm and 11 mm, respectively).

Conclusion: Within the limitations of this study, the Zeiss Extaro 300 was superior in terms of resolution and depth of field as compared to the other two DOMs.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7883792PMC
http://dx.doi.org/10.4103/JCD.JCD_191_20DOI Listing

Publication Analysis

Top Keywords

depth field
20
resolution depth
16
zeiss extaro
12
extaro 300
12
dental operating
8
compare resolution
8
edmund optics
8
optics barrington
8
louis usa
8
maximum resolution
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!