Purpose: In this study, two intraocular lenses (spherical IOL SA60AT and aspherical IOL SN60WF) are examined in an eye model under conditions of misalignment (defocus, decentration and tilt). The lenses are rated using the contrast sensitivity function (CSF) based on Barten's physical model. The square root integral (SQRI) method is used as a quality criterion comparable to the subjective image quality assessment of the human eye.
Methods: The IOLs to be tested are decentered from 0 to 1mm and tilted from -5 to +5 degrees in the Navarro eye model (optimized for far-point 6m and pupil aperture 3mm). The defocus of the IOLs is ±0.1mm at the anterior chamber depth (ACD). The optical modulation transfer function (MTF) is simulated with a ray tracing program. The SQRI is calculated using this MTF and the Barten CSF model (for in-focus at aperture 3 and 4.5mm and for defocus at 3mm).
Results: With increasing decentration, the spherical IOL shows a significantly smaller loss of quality for both apertures compared to the aspherical lens. With an aperture of 4.5mm, the image quality of the aspherical IOL is better for small decentration and tilt. The loss of quality of the spherical IOL increases with increasing tilt in both directions. In contrast, the image quality of the aspherical IOL is reduced under decentration for certain tilt values. For ACD-0.1mm, both IOLs behave similarly to the in-focus situation. For ACD+0.1mm, the influence of tilt without decentration is small for both IOLs. With increasing decentration, the quality loss of the aspherical IOL is similar to that in-focus and greater than that of the spherical lens.
Conclusion: In general, under the same conditions the spherical SA60AT displays a lower tolerance in loss of quality of subjective vision with lens alignment errors, in comparison to the aspherical SN60WF, limited by certain combinations of decentration and tilt according to this study. This study shows a way to evaluate IOLs based on the subjective visual performance of the eye.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.zemedi.2021.01.004 | DOI Listing |
Purpose: To observe and explore the correlation between visual outcomes and intraocular lens (IOL) stability after tri-focal IOL implantation in eyes with high myopia.
Methods: Patients with highly myopic cataract (axial length > 26 mm) were enrolled in this prospective study. Thirty-one eyes (31 patients) received implantation of a trifocal IOL (AcrySof IQ PanOptix TFNT00).
Retina
January 2025
National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.
Purpose: Anterior chamber air injection (ACAI) is a surgical technique used to decrease the occurrence of postoperative intraocular lens (IOL) haptic dislocation following phacovitrectomy with gas/air tamponade. The impact of this technique on IOL stability remains uncertain, prompting the design of this study to investigate further.
Methods: This study included 51 eyes of 51 patients who underwent phacovitrectomy with gas/air tamponade.
Sci Rep
December 2024
Visual Optics Lab Antwerp (VOLANTIS), Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
In cataract surgery, post-surgical stability of the intraocular lens plays a major role. This study aims to explore how the size and decentration of the capsulorhexis affect intraocular lens decentration and tilt by using numerical methods. Finite element models included zonules, ciliary body, capsular bag, and an IOL with two open-loop haptics were built.
View Article and Find Full Text PDFIntroduction: The aim of our study was to assess the outcome of Gore-Tex sutures in minimally invasive scleral fixation of subluxated posterior chamber intraocular lenses (PCIOLs) and to demonstrate a method for validating the lens position.
Methods: Retrospective study of patients who underwent lasso in-the-bag scleral fixation of a subluxated PCIOL using the snare technique with Gore-Tex suture from 2019 to 2021 in a single tertiary medical center. Functional outcome was analyzed by clinical assessment, and anatomical outcome, by ultrasound biomicroscopy (UBM).
Eur J Ophthalmol
December 2024
Department of Ophthalmology, ARNAS Civico Hospital, Palermo, Italy.
Background: To evaluate intrascleral plug stability and correct centring of the IOL in three different surgical techniques using intraoperative anterior segment OCT.
Methods: This retrospective scientific study was conducted by reviewing patient records from the Ophthalmology Unit at the Civico Hospital of Palermo, covering the period between 2021 and 2022. Three distinct groups of patients, each consisting of 7 eyes from 7 individual patients, were included in this study.