Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Confirmatory testing of SARS-CoV-2 results is essential to reduce false positives, but comes at a cost of significant extra workload for laboratories and increased turnaround time. A balance must be sought. We analysed our confirmatory testing pathway to produce a more refined approach in preparation for rising case numbers.
Methods: Over a 10-week low prevalence period we performed confirmatory testing on all newly positive results. Turnaround time was measured and results were analysed to identify a threshold that could be applied as a cut-off for future confirmatory testing and reduce overall workload for the laboratory.
Results: Between 22/06/20 and 31/08/20 confirmatory testing was performed on 108 newly positive samples, identifying 32 false positive results (30 %). Turnaround time doubled, increasing by an extra 17 h. There was a highly statistically significant difference between initial Relative Light Unit (RLU) of results that confirmed compared to those that did not, 1176 vs 721 (P < 0.00001). RLU = 1000 was identified as a suitable threshold for confirmatory testing in our laboratory: with RLU ≥ 1000, 55/56 (98 %) confirmed as positive, whereas with RLU < 1000 only 12/38 (32 %) confirmed.
Conclusions: False positive SARS-CoV-2 tests can be identified by confirmatory testing, yet this may significantly delay results. Establishing a threshold for confirmatory testing streamlines this process to focus only on samples where it is most required. We advise all laboratories to follow a similar process to identify thresholds that trigger confirmatory testing for their own assays, increasing accuracy while maintaining efficiency for when case numbers are high.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7877812 | PMC |
http://dx.doi.org/10.1016/j.jcv.2021.104762 | DOI Listing |