Purpose: The aim of this study is to evaluate the accuracy of three newer generation formulae (Barrett Universal II, EVO, Hill-RBF 2.0) for calculation of power of two standard IOLs-the Acrysof IQ and Tecnis ZCB00 across all axial lengths.

Methods: In this retrospective series, 206 eyes of 206 patients, operated for cataract surgery with above two IOLs over the last 6 months, were included in the study. Preoperative biometry measurements were obtained from LenstarLS900. By using recommended lens constants, the mean error for each formula was calculated and compared. Then, the optimized IOL constants were calculated to reduce the mean error to zero. Mean and median absolute errors were calculated for all eyes and separately for short (AL<22.5 mm), medium (22.5-24.5 mm), and long eyes (>24.5 mm). Absolute errors and percentages of eyes within prediction errors of ±0.25 D, ±0.50 D, ±0.75 D, and ±1.00 D were compared.

Results: Prediction error with using recommended lens constants was significantly lower in the Barrett Universal II formula as compared to the other two formulae. However, after optimizing lens constants, there were no significant differences in the absolute errors between the three formulae. The formulae ranked by mean absolute error were as follows: Barrett Universal II (0.304 D), EVO (0.317 D), and Hill-RBF (0.322) D. There were no significant differences between absolute errors in the three formulae in each of the short-, medium-, and long-eye subgroups.

Conclusion: With proper lens constant optimization, the Barrett Universal II, EVO, and Hill-RBF 2.0 formulae were equally accurate in predicting IOL power across the entire range of axial lengths.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7942088PMC
http://dx.doi.org/10.4103/ijo.IJO_943_20DOI Listing

Publication Analysis

Top Keywords

barrett universal
16
absolute errors
16
lens constants
12
three newer
8
newer generation
8
axial lengths
8
universal evo
8
evo hill-rbf
8
recommended lens
8
differences absolute
8

Similar Publications

Purpose: To apply a new method, the Taylor Diagram, and a new concept, the centered root mean square error (cRMSE), in evaluating the performance of IOL formulas.

Methods: The preoperative biometrics were measured with the IOLMaster 700 (Carl Zeiss Meditec) and the postoperative spherical equivalent refraction (SER) was calculated in 888 anonymous patients. The Taylor Diagram was applied to visualize the centered root mean square error (cRMSE) and the correlation coefficient between the predictions and the observations (R).

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcomes of a new FDA approved toric aspheric hydrophobic acrylic intraocular lens.

Patients And Methods: This is a single surgeon, single-arm, on-label, non-randomized, prospective observational study. Thirty patients underwent bilateral cataract surgery (60 eyes) with placement of a Clareon™ Toric IOL in each eye.

View Article and Find Full Text PDF

Objective: This study investigates the refractive accuracy of eight intraocular lens (IOL) power calculation formulas in patients with postoperative refractive surprise after phacoemulsification. It aims to determine if a different formula could result in better refractive outcomes in these eyes.

Methods And Analysis: We retrospectively reviewed consecutive patients undergoing uncomplicated phacoemulsification as a sole procedure between March 2007 and September 2020 at the University of Washington by glaucoma subspecialists as part of a study investigating cataract surgery in normal eyes.

View Article and Find Full Text PDF

Background: To evaluate the ocular biometry agreement and prediction of postoperative refractive outcomes obtained using two swept-source optical coherence tomography (SS-OCT) biometers: Anterion (Heidelberg Engineering, Heidelberg, Germany) and Argos (Alcon, Fort Worth, TX, USA).

Methods: Ambispective analysis was conducted on 105 eyes at the Samsung Medical Center, Seoul, Republic of Korea, between June 2021 and March 2022. Biometric values were assessed using both devices before cataract surgery.

View Article and Find Full Text PDF

Purpose: To evaluate the outcomes of scleral-fixated intraocular lenses (IOLs) implanted using either Yamane technique or Gore-Tex suture fixation, in comparison to intracapsular lens fixation, and to assess the efficacy of various lens formulas in achieving predicted refractive targets.

Patients And Methods: This study included 45 eyes from 44 patients with scleral-fixated IOLs, comprising 37 Yamane eyes and 8 Gore-Tex eyes. Preoperative refractive predictions from various formulae were compared with final postoperative refraction.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!