  • * Methods: Researchers reviewed 135 procedures performed on 116 patients, collecting data on demographics, symptoms, and various outcomes, including procedural success and complications related to the device and procedure.
  • * Results: The study found an overall procedural success rate of 87.4% with no device-related complications; however, some minor complications like access site hematoma (5.2%) and very low rates of major bleeding (0.7%) were reported, alongside 30-day adverse

Article Abstract

Objective: Our aim was to evaluate the efficacy and safety outcomes of the Pioneer Plus catheter (Philips, San Diego, Calif) and report the in-hospital and 30-day outcomes of lower extremity chronic total occlusion (CTO) interventions assisted by the Pioneer Plus catheter. In addition, we explored the factors associated with procedural success.

Methods: We conducted a retrospective review of 135 consecutive procedures in 116 patients from July 2011 to September 2018 performed by eight operators with various levels of experience at a high-volume center where the Pioneer Plus catheter was used for lower extremity CTO. The patient demographics, preprocedural symptoms, preprocedural testing results, procedural setting, and angiography findings were abstracted. The outcomes were divided into device-related and procedure-related outcomes. Device-related efficacy outcome included procedural success. Device-related safety outcomes included device-related complications. Procedure-related outcomes included procedure-related complications, 30-day major adverse cardiovascular events, and 30-day major adverse limb events. We conducted univariate comparisons of the provider, patient, and procedural characteristics stratified by procedural success.

Results: Procedural success was observed in 118 procedures overall (87.4%), and success rates ≤95.8% were observed for operators with an experience level of >25 devices deployed. No device-related complications, such as pseudoaneurysm formation, vessel perforation, or arteriovenous fistula formation, were observed. The Pioneer Plus catheter was mostly often used for CTO in the superficial femoral and popliteal arteries. Overall, the procedure-related complications included access site hematoma (5.2%), major bleeding (0.7%), pseudoaneurysm formation (0.7%), distal embolization (1.5%), and acute arterial thrombosis (1.5%). The 30-day major adverse limb events included index limb unplanned amputation (0.7%), index limb reintervention (4.4%), and index limb acute limb ischemia (0.7%) and occurred in 5.9% of the procedures. The only factor associated with procedural success was operator experience (P < .0001).

Conclusions: The results from the present study have shown that Pioneer Plus catheter use is safe and effective when used to cross lower extremity CTO. However, further investigation is needed to identify patient- and provider-level factors to optimize patient outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2021.01.045DOI Listing

Publication Analysis

Top Keywords

pioneer catheter
24
lower extremity
16
procedural success
12
30-day major
12
major adverse
12
outcomes
8
outcomes pioneer
8
extremity chronic
8
chronic total
8
safety outcomes
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!