Background: Total hip arthroplasty with ceramic-on-ceramic articulation aims to decrease wear, osteolysis, and aseptic loosening. A metal-backed ceramic liner was developed to reduce the risk of liner fracture. However, a significant number of cases of mal-seating of the metal-backed ceramic liner were observed in the 2000s, and there were concerns about their outcome. This review aims to investigate the long-term performance of the mal-seated ceramic liner.

Methods: From July 2003 to March 2007, 35 ceramic-on-ceramic total hip arthroplasties were performed with the Trident acetabular system. Clinical assessment, radiological analysis, and outcome assessment were performed. The prevalence of liner mal-seating and its long-term outcomes were investigated.

Results: There was liner mal-seating in 8 hips (22.9%). One liner was exchanged in the early postoperative period. No revision surgery was required for the remaining 7 hips at a mean follow-up of 14 years. All patients were free of hip pain with a mean Harris Hip Score of 94.7 at the most recent follow-up. No adverse event was observed.

Conclusion: The long-term outcomes of the mal-seated liner were favorable. However, surgeons should exercise meticulous surgical technique to achieve a properly aligned liner within the acetabular shell to minimize this potentially correctable error.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2021.01.024DOI Listing

Publication Analysis

Top Keywords

total hip
12
ceramic-on-ceramic articulation
8
hip arthroplasty
8
metal-backed ceramic
8
liner
8
ceramic liner
8
liner mal-seating
8
long-term outcomes
8
hip
5
long-term
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!