Background: The objective is to compare, by the means of finite elements analysis, the biomechanical behavior of a conventional stem of proven performance with a short stem based on the same fixation principles.

Methods: A 3D femur was modeled from CT scan data, and real bone density measures were incorporated into it. Load stresses were applied to that bone in 3 different scenarios: without prosthesis, with the conventional stem, and with the short stem. Different bone loading patterns were compared by Gruen's zones both visually and statistically using Welch's test.

Results: The implantation of a stem generates a certain degree of stress shielding in the surrounding bone, but the pattern of the change is very similar in the compared stem models. Although there is statistical significance ( < 0.01) in the mean stress variation in most of the Gruen's zones, the magnitude of the difference is always under 2 MPa (range: 0.01 - 1.74 MPa).

Conclusions: The bone loading patterns of the traditional stem and the short stem are very similar. Although there is no evidence of a link between biomechanics and clinical outcomes, our results may suggest that theoretical advantages of short stems can be exploited without the fear of altering bone loading patterns.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7856394PMC
http://dx.doi.org/10.1016/j.artd.2020.11.022DOI Listing

Publication Analysis

Top Keywords

stem short
12
short stem
12
bone loading
12
loading patterns
12
stem
9
biomechanical behavior
8
conventional stem
8
gruen's zones
8
bone
6
short
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!