Relationship of Antenna Work and Ablation Cavity Volume Following Percutaneous Microwave Ablation of Hepatic Tumors.

J Vasc Interv Radiol

Department of Vascular and Interventional Radiology, University of Virginia, Charlottesville, Virginia.

Published: April 2021

Purpose: To formulate a statistical model relating ablation time, power, and work with posttreatment cavity volume following percutaneous microwave ablation of hepatic tumors in vivo.

Materials And Methods: A retrospective review (October 2015 to October 2018) yielded 122 hepatic tumors treated with microwave ablation. Ablation cavity dimensions were measured at 1-month follow-up examination and calculated using an ellipsoid volume formula. The antenna manufacturer (Neuwave Medical, Madison, Wisconsin) provided the activation time and energy used to calculate the antenna work. Generalized estimating equations with ordinary least-squares regression models were obtained to relate tumor volume with cumulative antenna work. Coefficient of determination (R) and mean square error were used as statistical measures of model prediction performance.

Results: There is a logarithmic relationship between postablation cavity volume (cm) and cumulative work (kJ), represented by the formula: log cm = -0.4583 + 0.9887 × cumulative work (log10 kJ) (R = 0.41, mean square error, 0.102). Ablation volumes were predicted as a function of antenna work, calculated using an antilog transformation. When a single antenna was used, ablation cavity volume was predicted using a generalized estimating equation ordinary least-squares regression model of power and time: logcm= -0.0546 + 0.0485 × total time (min) + 0.0107 × power (W) (R = 0.30; mean square error, 0.106). Using this model, a nomogram was developed to predict the postablation cavity volume based on total activation time and target power.

Conclusion: There is a logarithmic relationship between the ablation work and posttreatment ablation cavity volume, which can be expressed in a nomogram when using a single probe.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvir.2020.12.012DOI Listing

Publication Analysis

Top Keywords

cavity volume
24
antenna work
16
ablation cavity
16
microwave ablation
12
hepatic tumors
12
square error
12
ablation
10
work
8
volume
8
volume percutaneous
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!