Background: One of the most prevalent causes of fetal hypoxia leading to stillbirth is placental insufficiency. Hemodynamic changes evaluated with Doppler ultrasound have been used as a surrogate marker of fetal hypoxia. However, Doppler evaluation cannot be performed continuously. As a first step, the present work aimed to evaluate the performance of miniaturized electrochemical sensors in the continuous monitoring of oxygen and pH changes in a model of acute hypoxia-acidosis.

Methods: pH and oxygen electrochemical sensors were evaluated in a ventilatory hypoxia rabbit model. The ventilator hypoxia protocol included 3 differential phases: basal (100% FiO), the hypoxia-acidosis period (10% FiO) and recovery (100% FiO). Sensors were tested in blood tissue (ex vivo sensing) and in muscular tissue (in vivo sensing). pH electrochemical and oxygen sensors were evaluated on the day of insertion (short-term evaluation) and pH electrochemical sensors were also tested after 5 days of insertion (long-term evaluation). pH and oxygen sensing were registered throughout the ventilatory hypoxia protocol (basal, hypoxia-acidosis, and recovery) and were compared with blood gas metabolites results from carotid artery catheterization (obtained with the EPOC blood analyzer). Finally, histological assessment was performed on the sensor insertion site. One-way ANOVA was used for the analysis of the evolution of acid-based metabolites and electrochemical sensor signaling results; a t-test was used for pre- and post-calibration analyses; and chi-square analyses for categorical variables.

Results: At the short-term evaluation, both the pH and oxygen electrochemical sensors distinguished the basal and hypoxia-acidosis periods in both the in vivo and ex vivo sensing. However, only the ex vivo sensing detected the recovery period. In the long-term evaluation, the pH electrochemical sensor signal seemed to lose sensibility. Finally, histological assessment revealed no signs of alteration on the day of evaluation (short-term), whereas in the long-term evaluation a sub-acute inflammatory reaction adjacent to the implantation site was detected.

Conclusions: Miniaturized electrochemical sensors represent a new generation of tools for the continuous monitoring of hypoxia-acidosis, which is especially indicated in high-risk pregnancies. Further studies including more tissue-compatible material would be required in order to improve long-term electrochemical sensing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7863274PMC
http://dx.doi.org/10.1186/s12967-021-02715-7DOI Listing

Publication Analysis

Top Keywords

electrochemical sensors
24
vivo sensing
16
miniaturized electrochemical
12
long-term evaluation
12
electrochemical
10
monitoring oxygen
8
sensors
8
model acute
8
fetal hypoxia
8
continuous monitoring
8

Similar Publications

The ultrahigh-sensitive detection of HS is reported using a novel dual-ligand metal-organic framework (MOF) electrochemiluminescence (ECL) sensor. By combining tetrakis(4-carboxyphenyl) porphyrin (TCPP) and 1,3,6,8-tetrakis(4-carboxyphenyl) pyrene (TBAPy) as ligands and employing zirconium as the metal source, a spindle-shaped Zr-PyTCPPMOF was successfully designed and synthesized. Notably, the multiple nitrogen structures of porphyrin provided abundant binding sites for sulfur (S), further enhancing the ECL signal of Zr-PyTCPPMOF.

View Article and Find Full Text PDF

Carboxymethyl hexanoyl chitosan drop-coated simple disposable paper electrochemical sensor for quality monitoring of vanillin.

Int J Biol Macromol

December 2024

Nanomaterial research laboratory (NMRL), Smart Materials And Devices, Yenepoya Research Centre, Yenepoya (Deemed to be university), Deralakatte, Mangalore 575 018, India; Centre for Nutrition Studies, Yenepoya (Deemed to be University), Deralakatte, Mangalore 575 018, India. Electronic address:

The food and pharmaceutical sectors frequently utilize vanillin (VAN), a food ingredient with a pleasing flavor and aroma. However, excessive consumption of VAN causes several health problems, including liver and kidney damage, headaches, skin conditions, nausea, and vomiting. To prevent health problems, it is crucial to identify and control the amount of VAN in food and drugs.

View Article and Find Full Text PDF

Highly sensitive electrochemical sensor for lead ions based on Bi-MOF/conducting polymer composites.

Chemosphere

December 2024

Frontiers Science Center for Deep Ocean Multispheres and Earth System, and Key Laboratory of Marine Chemistry Theory and Technology, Ministry of Education, Ocean University of China, Qingdao 266100, China. Electronic address:

Herein, conductive polyaniline (PANI) was chemically polymerized on the surface of a bismuth-based metal-organic framework (Bi-MOF) to form conductive PANI@Bi-MOF composites. FT-IR and PXRD measurements verified the successful production of PANI@Bi-MOF, whereas SEM, TEM, and EDAX mapping demonstrated that PANI was uniformly coated on the surface of Bi-MOF. The resulting PANI@Bi-MOF composites were characterized by cyclic voltammetry (CV and electrochemical impedance spectroscopy (EIS), then used to develop a sensitive electrochemical sensor for the detection of lead ions based on differential pulse anodic stripping voltammetry (DPASV).

View Article and Find Full Text PDF

Metol (MTO), a commonly used photographic developer, has become an environmental pollutant due to its extensive use and subsequent release into water sources. The accumulation of MTO poses significant risks, including aquatic toxicity and potential bioaccumulation, leading to adverse effects on ecosystems. To address these environmental challenges, we developed a La₂NiO combined with graphene oxide (La₂NiO₄@GO) nanocomposite modified glassy carbon electrode (GCE) for the ultrasensitive detection of MTO.

View Article and Find Full Text PDF

Wearable multiple sensing platform for enhanced biomolecules monitoring in food.

Food Chem

December 2024

College of Chemistry and Chemical Engineering, Engineering Research Center of Dairy Quality and Safety Control Technology, Ministry of Education, Inner Mongolia University, 235 University West Street, Hohhot, China. Electronic address:

Monitoring of biomolecules in food plays a crucial role in safeguarding human health. Prevalent biomolecule monitoring systems are constructed predominantly from rigid materials and have inherent limitations in detection capabilities. Wearable sensors have increasingly captured attention, significantly propelling the evolution of biomolecular detection process.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!