Purpose: To examine the contribution of anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW) measurements to intraocular lens (IOL) power calculations using the Barrett Universal II (BUII) formula.

Methods: Measurements taken with the IOLMaster 700 (Carl Zeiss, Meditec AG, Jena, Germany) swept-source biometry of 501 right eyes of 501 consecutive patients undergoing cataract extraction surgery between January 2019 and March 2020 were reviewed. IOL power was calculated using the BUII formula, first through the inclusion of all measured variables and then by using partial biometry data. For each calculation method, the IOL power targeting emmetropia was recorded and compared for the whole cohort and stratified by axial length (AL) of the measured eye.

Results: The mean IOL power calculated for the entire cohort using all available parameters was 19.50 ± 5.11 diopters (D). When comparing it to the results obtained by partial biometry data, the mean absolute difference ranged from 0.05 to 0.14 D; < 0.001. The optional variables (ACD, LT, WTW) had the least effect in long eyes (AL ≥ 26 mm; mean absolute difference ranging from 0.02 to 0.07 D; < 0.001), while the greatest effect in short eyes (AL ≤ 22 mm; mean absolute difference from 0.10 to 0.21 D; < 0.001). The percentage of eyes with a mean absolute IOL dioptric power difference more than 0.25 D was the highest (32.0%) among the short AL group when using AL and keratometry values only.

Conclusions: Using partial biometry data, the BUII formula in small eyes (AL ≤ 22 mm) resulted in a clinically significant difference in the calculated IOL power compared to the full biometry data. In contrast, the contribution of the optional parameters to the calculated IOL power was of little clinical importance in eyes with AL longer than 22 mm.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7867297PMC
http://dx.doi.org/10.3390/jcm10030542DOI Listing

Publication Analysis

Top Keywords

iol power
24
biometry data
16
partial biometry
12
absolute difference
12
barrett universal
8
power calculated
8
buii formula
8
eyes ≤
8
calculated iol
8
iol
7

Similar Publications

Purpose: To investigate different measures for corneal astigmatism in the context of reconstructed corneal astigmatism (recCP) as required to correct the pseudophakic eye, and to derive prediction models to map measured corneal astigmatism to recCP.

Methods: Retrospective single centre study of 509 eyes of 509 cataract patients with monofocal (MX60P) IOL. Corneal power measured with the IOLMaster 700 keratometry (IOLMK), and Galilei G4 keratometry (GK), total corneal power (TCP2), and Alpin's integrated front (CorT) and total corneal power (CorTTP).

View Article and Find Full Text PDF

Objectives: This study compared the visual outcomes and optical quality of two monofocal, aspheric intraocular lenses (IOLs; CT LUCIA 621P, Carl Zeiss Meditec; Eyhance ICB00, Johnson & Johnson Vision) by evaluating visual acuity, contrast sensitivity, and higher-order aberrations 1 month post-cataract surgery.

Methods: In this retrospective, comparative study, 120 eyes (72 patients) that underwent cataract surgery with either CT LUCIA 621P (Lucia group) or Eyhance ICB00 (Eyhance group) implantation (60 eyes/group) were retrospectively investigated. Visual acuity at various distances and defocus curves were measured 1 month postoperatively.

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcomes of a new FDA approved toric aspheric hydrophobic acrylic intraocular lens.

Patients And Methods: This is a single surgeon, single-arm, on-label, non-randomized, prospective observational study. Thirty patients underwent bilateral cataract surgery (60 eyes) with placement of a Clareon™ Toric IOL in each eye.

View Article and Find Full Text PDF

Objective: This study investigates the refractive accuracy of eight intraocular lens (IOL) power calculation formulas in patients with postoperative refractive surprise after phacoemulsification. It aims to determine if a different formula could result in better refractive outcomes in these eyes.

Methods And Analysis: We retrospectively reviewed consecutive patients undergoing uncomplicated phacoemulsification as a sole procedure between March 2007 and September 2020 at the University of Washington by glaucoma subspecialists as part of a study investigating cataract surgery in normal eyes.

View Article and Find Full Text PDF

Purpose: This study evaluates the discrepancies between ACV measurements obtained from the Heidelberg Anterion and Zeiss IOLMaster 700 and investigates the significance of ACV and other ocular biometry parameters.

Patients And Methods: To investigate intraocular fluid circulation, a robust formula was developed for ACV measurement using the Zeiss IOLMaster 700. A pilot study was conducted to validate this formula, which relied on WTW, CCT, and ACD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!