Objective: The aim of this study was to establish the frequency of cutaneous complications due to surgically implanted bone conduction hearing aids in recipients treated at a tertiary otolaryngology center. Additionally, based on the literature review, we propose a scale to standardize the report of cutaneous reactions related to transcutaneous systems to objectify future studies.

Methods: A retrospective chart review was conducted for patients who were implanted with bone conduction hearing systems between 2004 and 2018. Patients with minimum follow up of 6 months were included. Patient demographics, surgical characteristics and clinical course was analyzed by number and skin reactions. Transcutaneous and percutaneous systems were analyzed separately. The study was approved by the local IRB.

Results: A total of 104 devices were implanted on 88 patients with an average follow up of 4.04 (range of 0.5 - 10 years). Out of the total of patients, 49 (55.7%) developed at least one episode of inflammatory/infectious skin reaction at surgical incision site. A total of 148 episodes of local infections during the entire follow-up period were registered, mostly mild in severity with no triggering factors identified. The majority of the initial episodes of infection occurred within the first 3 years of follow-up. Out of the total of patients, 47 (53.4%) reported pain at the surgical site at some point throughout follow-up, not associated with clinically evident infection.

Conclusion: The incidence of skin complications in our series seemed higher compared to previous reports. Minor complications were the most common and responded well to topical treatment. No triggering factors were identified as the cause of the infections, nor to explain the frequency or the severity of such an adverse reaction. Isolated pain was present in the majority of patients with conductive hearing devices, even without signs of active infection. Due to its high incidence it should be assessed in all patients that receive an implant. Based on a search of the literature, a scale to standardize cutaneous complications of transcutaneous implants was undertaken, but further studies are needed to validate such a scale.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.anl.2021.01.020DOI Listing

Publication Analysis

Top Keywords

cutaneous complications
12
bone conduction
12
conduction hearing
12
implanted bone
8
scale standardize
8
reactions transcutaneous
8
total patients
8
triggering factors
8
factors identified
8
patients
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!