Background: Re-entry devices are used regularly in subintimal recanalization of chronic occlusions of the iliac and femoro-popliteal arteries and significantly contribute to the high success rate of these interventions. However, the use in tibio-peroneal arteries has only been described in few cases so far. The present work is a retrospective evaluation of the Outback® re-entry device for gaining targeted true lumen access at the level of the tibio-peroneal arteries.
Methods: From 9/2017 until 10/2020 the Outback® catheter was used in case of failed spontaneous re-entry at the level of the tibio-peroneal arteries in 14 patients either instead of the usual retrograde approach via a pedal/distal-crural access (n = 11) or in combination with it (n = 3). Baseline demographic and clinical data, morphologic characteristics of the occlusions, procedural succedss, as well as the Society of Vascular Surgery (SVS) runoff score before and after revascularization were documented.
Results: All patients (median age: 78 years; range: 66-93) suffered from peripheral artery occlusive disease Rutherford stage 4 to 6 with a median lesion length of 12 cm (range: 7-35). Technical and procedural success was achieved in all 14 patients. The mean re-entry accuracy was 0.25 cm (range: 0-0.8). The SVS runoff score improved from a median of 14.5 (interquartile range IQR: 10.8-16.4) to 7 (IQR: 6.3-7) (p < 0.01).
Conclusions: The use of the Outback® catheter for targeted tibio-peroneal re-entry is associated with a high technical and procedural success rate and should be considered in case of otherwise failed ante- and retrograde recanalization.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7843800 | PMC |
http://dx.doi.org/10.1186/s42155-021-00206-w | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!