Article Abstract

Background: The leucocyte esterase (LE) strip test often is used to diagnose periprosthetic joint infection (PJI). In accordance with the manufacturer's directions, the LE strip test result is read 3 minutes after exposing it to joint fluid, but this has not been supported by robust research. Moreover, we have noted that the results of the LE strip test might change over time, and our previous studies have found that centrifugation causes the results of the LE strip test to degrade. Still, there is no evidence-based recommendation as to when to read the LE strip test to maximize diagnostic accuracy, in general, and the best reading times for the LE strip test before and after centrifugation need to be determined separately, in particular.

Questions/purposes: (1) What is the optimal timing for reading LE strip test results before centrifugation to diagnose PJI? (2) What is the optimal timing for reading LE strip test results after centrifugation to diagnose PJI?

Methods: This study was a prospective diagnostic trial. In all, 120 patients who were scheduled for revision arthroplasty and had signs of infection underwent joint aspiration in the outpatient operating room between July 2018 and July 2019 and were enrolled in this single-center study. For inclusion, patients must have had a diagnosis of PJI or nonPJI, valid synovial fluid samples, and must not have received antibiotics within 2 weeks before arthrocentesis. As such, 36 patients were excluded; 84 patients were included for analysis, and all 84 patients agreed to participate. The 2018 International Consensus Meeting Criteria (ICM 2018) was used for the classification of 49 patients with PJI (score ≥ 6) and 35 without PJI (score ≤ 2). The classification was used as the standard against which the different timings for reading LE strips were compared. All patients without PJI were followed for more than 1 year, during which they did not report the occurrence of PJI. All patients were graded against the diagnostic criteria regardless of their LE strip test results. In 83 patients, one drop of synovial fluid (50 μL) was applied to LE strips before and after centrifugation, and in one patient (without PJI), the sample was not centrifuged because the sample volume was less than 1.5 mL. The results of the strip test were read on an automated colorimeter. Starting from 1 minute after centrifugation, these strips were automatically read once every minute, 15 times (over a period of 16 minutes), and the results were independently recorded by two observers. Results were rated as negative, ±, 1+, and 2+ upon the machine reading. Grade 2+ (dark purple) was used as the threshold for a positive result. An investigator who was blinded to the study performed the statistics. Optimal timing for reading the LE strip before and after centrifugation was determined by using receiver operative characteristic (ROC) analysis. The specificity, sensitivity, and positive predictive and negative predictive values were calculated for key timepoints.

Results: Before centrifugation, the area under the curve was the highest when the results were read at 5 minutes (0.90 [95% CI 0.83 to 0.98]; sensitivity 0.88 [95% CI 0.75 to 0.95]; specificity 0.89 [95% CI 0.72 to 0.96]). After centrifugation, the area under the curve was the highest when the results were read at 10 minutes (0.92 [95% CI 0.86 to 0.98]; sensitivity 0.65 [95% CI 0.50 to 0.78]; specificity 0.97 [95% CI 0.83 to 1.00]).

Conclusion: The LE strip test results are affected by time and centrifugation. For samples without centrifugation, we found that 5 minutes after application was the best time to read LE strips. We cannot deny the use of centrifuges because this is an effective way to solve the sample-mingling problem at present. We recommend 10 minutes postapplication as the most appropriate time to read LE strips after centrifugation. Multicenter and large-sample size studies are warranted to further verify our conclusion.

Level Of Evidence: Level II, diagnostic study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8133206PMC
http://dx.doi.org/10.1097/CORR.0000000000001609DOI Listing

Publication Analysis

Top Keywords

strip test
44
optimal timing
16
timing reading
16
strip
13
read minutes
12
centrifugation
12
test centrifugation
12
reading strip
12
test
11
patients
9

Similar Publications

Paper-Based Aptasensor Assay for Detection of Food Adulterant Sildenafil.

Biosensors (Basel)

December 2024

Department of Medical Biology, School of Medicine, Atilim University, Ankara 06830, Turkey.

Sildenafil is used to treat erectile dysfunction and pulmonary arterial hypertension but is often illicitly added to energy drinks and chocolates. This study introduces a lateral flow strip test using aptamers specific to sildenafil for detecting its illegal presence in food. The process involved using graphene oxide SELEX to identify high-affinity aptamers, which were then converted into molecular gate structures on mesoporous silica nanoparticles, creating a unique signaling system.

View Article and Find Full Text PDF

Gatifloxacin (GAT), an antibiotic belonging to the fluoroquinolone (FQ) class, is a toxicant that may contaminate food products. In this study, a method of ultrasensitive immunochromatographic detection of GAT was developed for the first time. An indirect format of the lateral flow immunoassay (LFIA) was performed.

View Article and Find Full Text PDF

Nucleic Acid Lateral Flow Assay Implemented with Isothermal Gene Amplification of SARS-CoV-2 RNA.

Biosensors (Basel)

December 2024

Department of Bioscience and Biotechnology, Konkuk University, 120 Neungdong-ro, Gwangjin-gu, Seoul 05029, Republic of Korea.

We developed a rapid and sensitive diagnostic platform that integrates isothermal viral gene amplification with a nucleic acid lateral flow assay (NALFA) to detect SARS-CoV-2 RNA. Isothermal gene amplification was performed by combining reverse transcription of viral RNA with recombinase polymerase amplification (RPA). In our diagnostic platform, DNA primers for the RPA reaction were modified by appending DNA tails, enabling the synthesis of tailed amplicon DNAs.

View Article and Find Full Text PDF

Background: Schistosoma spp. and other intestinal parasites are common in Ethiopia. During pregnancy, SCH increases the risk of adverse birth outcomes.

View Article and Find Full Text PDF

Drug checking services (DCS) are entities that allow recreational drug users to have street drug samples analyzed. Diverse analytical methods are applied for DCS, ranging from test strips to mass spectrometry (MS). This work evaluates the performance and utility of common methodologies used for DCS operating with off-site drug testing, while additionally assessing the potential of gas chromatography coupled to vapor phase infrared spectroscopy (GC-IR).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!