Background: Novel, portable blood gas analyzers (BGAs) may serve as essential point-of-care tools in remote regions, during air travel or in ambulance services but they have not been extensively validated.

Research Question: We compared accuracy of a portable BGA to a validated stationary device.

Methods: In healthy individuals and patients with chronic obstructive pulmonary disease participating in clinical field studies at different altitudes, arterial blood samples were obtained at rest and during exercise in a hospital at 760 m and in a high altitude clinic at 3100 m. Paired measurements by a portable BGA (EPOC, Siemens Healthcare) and a stationary BGA (Rapidpoint500, Siemens Healthcare) were performed to compute bias (mean difference) and limits of agreement (95% CI of bias).

Results: Of 105 individuals, 248 arterial blood samples were analyzed, 108 at 760 m, 140 at 3100 m. Ranges of values measured by portable BGA were: pH 7.241-7.473, PaCO 21.5-52.5 mmHg, and PaO 45.5-107.1 mmHg. Bias (95% CI) between devices were: pH 0.007 (-0.029 to 0.044), PaCO -0.3 mmHg (-4.8 to 4.2), and PaO -0.2 mmHg (-9.1 to 4.7). For pH, agreement between devices was improved by the equation to correct pH by portable BGA = -1.37 + pH × 1.19; bias after correction -0.007 (-0.023 to 0.009). The portable BGA was easily handled and worked reliably.

Interpretation: Accuracy of blood gas analysis by the portable BGA in comparison to the reference BGA was adequate for clinical use. Because of portability and ease of handling, portable BGA are valuable diagnostic tools for use in everyday practice as well as under challenging field conditions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7820720PMC
http://dx.doi.org/10.3389/fphys.2020.600551DOI Listing

Publication Analysis

Top Keywords

portable bga
28
blood gas
12
bga
9
portable blood
8
challenging field
8
field conditions
8
high altitude
8
portable
8
arterial blood
8
blood samples
8

Similar Publications

Background: In recent years, there has been a significant increase in the number of patients requiring out-of-hospital intensive care. Although many of these patients are invasively ventilated, pulmonological care is often lacking. Additionally, up to 60-70% of these patients are suspected to have further potential for weaning.

View Article and Find Full Text PDF

Background: Novel, portable blood gas analyzers (BGAs) may serve as essential point-of-care tools in remote regions, during air travel or in ambulance services but they have not been extensively validated.

Research Question: We compared accuracy of a portable BGA to a validated stationary device.

Methods: In healthy individuals and patients with chronic obstructive pulmonary disease participating in clinical field studies at different altitudes, arterial blood samples were obtained at rest and during exercise in a hospital at 760 m and in a high altitude clinic at 3100 m.

View Article and Find Full Text PDF

Development of the Brief Geriatric Assessment for the General Practitioner.

J Nutr Health Aging

April 2021

Dr. Tzyy-Guey Tseng, No.100, Shin-Chuan 1st Road, Sanmin Dist., Kaohsiung City 80708, Taiwan (R.O.C.), Department of Family Medicine, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung, Taiwan, Email: Tel: +886-7-3121101 ext. 7021; Prof. Meng-Chih Lee, No. 199, Sec. 1, Sanmin Rd., West Dist., Taichung City 403, Taiwan (R.O.C.), Department of Family Medicine, Taichung Hospital, Ministry of Health and Welfare, Taiwan, Email: Tel: +886-4-22294411.

Objectives: The study aimed to develop a brief geriatric assessment (BGA) tool for the general practitioner to evaluate geriatric syndromes in community-dwelling older adults.

Design: A cross-sectional study.

Setting: 58 communities from four aging cities in Taiwan.

View Article and Find Full Text PDF

Ancestry informative single nucleotide polymorphisms (SNPs) can identify biogeographic ancestry (BGA); however, population substructure and relatively recent admixture can make differentiation difficult in heterogeneous Hispanic populations. Utilizing unrelated individuals from the Genomic Origins and Admixture in Latinos dataset (GOAL, n = 160), we designed an 80 SNP panel (Setser80) that accurately depicts BGA through STRUCTURE and PCA. We compared our Setser80 to the Seldin and Kidd panels via resampling simulations, which models data based on allele frequencies.

View Article and Find Full Text PDF

We analyzed 37 satellite reflectance algorithms and 321 variants for five satellites for estimating turbidity in a freshwater inland lake in Ohio using coincident real hyperspectral aircraft imagery converted to relative reflectance and dense coincident surface observations. This study is part of an effort to develop simple proxies for turbidity and algal blooms and to evaluate their performance and portability between satellite imagers for regional operational turbidity and algal bloom monitoring. Turbidity algorithms were then applied to synthetic satellite images and compared to in situ measurements of turbidity, chlorophyll-a (Chl-a), total suspended solids (TSS) and phycocyanin as an indicator of cyanobacterial/blue green algal (BGA) abundance.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!