Objectives: This multicentric study was carried out to investigate the impact of small field output factors (OFs) inaccuracies on the calculated dose in volumetric arctherapy (VMAT) radiosurgery brain plans.

Methods: Nine centres, realised the same five VMAT plans with common planning rules and their specific clinical equipment Linac/treatment planning system commissioned with their OFs measured values (OFbaseline). In order to simulate OFs errors, two new OFs sets were generated for each centre by changing only the OFs values of the smallest field sizes (from 3.2 × 3.2 cm to 1 × 1 cm) with well-defined amounts (positive and negative). Consequently, two virtual machines for each centre were recommissioned using the new OFs and the percentage dose differences ΔD (%) between the baseline plans and the same plans recalculated using the incremented (OFup) and decremented (OFdown) values were evaluated. The ΔD (%) were analysed in terms of planning target volume (PTV) coverage and organs at risk (OARs) sparing at selected dose/volume points.

Results: The plans recalculated with OFdown sets resulted in higher variation of doses than baseline within 1.6 and 3.4% to PTVs and OARs respectively; while the plans with OFup sets resulted in lower variation within 1.3% to both PTVs and OARs. Our analysis highlights that OFs variations affect calculated dose depending on the algorithm and on the delivery mode (field jaw/MLC-defined). The Monte Carlo (MC) algorithm resulted significantly more sensitive to OFs variations than all of the other algorithms.

Conclusion: The aim of our study was to evaluate how small fields OFs inaccuracies can affect the dose calculation in VMAT brain radiosurgery treatments plans. It was observed that simulated OFs errors, return dosimetric calculation accuracies within the 3% between concurrent plans analysed in terms of percentage dose differences at selected dose/volume points of the PTV coverage and OARs sparing.

Advances In Knowledge: First multicentre study involving different Planning/Linacs about undetectable errors in commissioning output factor for small fields.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8011246PMC
http://dx.doi.org/10.1259/bjr.20201354DOI Listing

Publication Analysis

Top Keywords

calculated dose
12
ofs
10
small field
8
field output
8
output factors
8
plans
8
vmat plans
8
multicentre study
8
ofs inaccuracies
8
ofs errors
8

Similar Publications

Association of pan-immune-inflammatory value with metabolic dysfunction-associated steatotic liver disease: findings from NHANES 2017-2020.

BMC Gastroenterol

January 2025

Department of Gastroenterology and Hepatology, The First Affiliated Hospital of Xiamen University, School of Medicine, Xiamen University, No. 55 Zhenhai Road, Xiamen, 361003, China.

Background: Metabolic dysfunction-associated steatotic liver disease (MASLD) has become the most common chronic liver disease worldwide. The pan-immune-inflammation value (PIV) has been proposed as a biomarker for assessing immune status and inflammation. There is currently no evidence regarding the effect of PIV on the risk of MASLD.

View Article and Find Full Text PDF

Background: Patients on a kidney transplant waiting list with antibodies against more than 80% of a panel reactive antibody (PRA) are difficult to transplant, even with national or regional programs. Desensitization treatment with high-dose intravenous immunoglobulin and rituximab could be offered to patients with a long waiting time for a cadaveric donor to improve their odds of finding a kidney.

Methods: This was a retrospective, single-center study including all hyperimmunized patients on the waiting list for a cadaveric kidney donor who received a desensitization treatment between 2010 and 2020.

View Article and Find Full Text PDF

Objective: To evaluate the added value of dynamic contrast-enhanced ultrasound (DCE-US) analysis in pre-operative differential diagnosis of small (≤20 mm) solid pancreatic lesions (SPLs).

Methods: In this retrospective study, patients with biopsy or surgerical resection and histopathologically confirmed small (≤20 mm) SPLs were included. One wk before biopsy/surgery, pre-operative B-mode ultrasound and contrast-enhanced ultrasound were performed.

View Article and Find Full Text PDF

Objectives: The purpose of this study was to investigate the fundamental properties of spot-scanning proton beams and compare them to Monte Carlo (MC) simulations, both with and without CT calibration, using spatially diverse combinations of materials.

Methods: A heterogeneous phantom was created by spatially distributing titanium, wax, and thermocol to generate six scenarios of heterogeneous combinations. Proton pencil beams ranging in energy from 100 to 226.

View Article and Find Full Text PDF

The tumorigenic dose 50 (TD) is a widely used measure of carcinogenic potency which has historically been used to determine acceptable intake limits for carcinogenic compounds. Although broadly used, the TD model was not designed to account for important biological factors such as DNA repair and cell compensatory mechanisms, changes in absorption, etc., leading to the development of benchmark dose (BMD) approaches, which use more flexible dose-response models that are better able to account for these processes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!