Objectives: High-frequency 10 kHz spinal cord stimulation (10 kHz-SCS) has achieved analgesia superior to traditional SCS in a number of studies. However, there is concern regarding long-term outcomes of 10 kHz-SCS. Prior work has suggested that explant rates are higher with 10 kHz-SCS. Our primary objective was to determine the explant rate of 10 kHz-SCS in a large patient cohort from multiple centers followed for at least 12 months after implant surgery.

Materials And Methods: We performed a retrospective chart review of all patients who received a 10 kHz-SCS implant before July 1, 2019. We abstracted patient demographics, implant date, primary site of pain, implant indication, explant date, and reason for explant. A total of 744 patients were included in the study analysis.

Results: Average age of the overall cohort was 65.53 years and 407 (54.7%) were women. Average follow-up for all patients was 793 days. There were a total of 76 explants (10.2%). The most common reason for explant was loss of efficacy, which accounted for 39 explants (51.3% of total explants, 5.2% of overall cohort). Female sex and radiculopathy as the SCS indication were associated with statistically significant decreased risk of 10 kHz-SCS explant.

Conclusions: We found 10 kHz-SCS explant rates to be similar to prior reported explant rates for traditional SCS devices. Patient-related factors including female sex and radiculopathy as the primary SCS indication may be protective factors against explantation.

Download full-text PDF

Source
http://dx.doi.org/10.1111/ner.13359DOI Listing

Publication Analysis

Top Keywords

explant rates
12
10 khz spinal
8
spinal cord
8
cord stimulation
8
744 patients
8
traditional scs
8
reason explant
8
total explants
8
female sex
8
sex radiculopathy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!