Interpretation of Near-Infrared Spectroscopy (NIRS) Signals in Skeletal Muscle.

J Funct Morphol Kinesiol

Department of Kinesiology, University of Georgia, Athens, GA 30602, USA.

Published: May 2019

Unlabelled: Near-infrared spectroscopy (NIRS) uses the relative absorption of light at 850 and 760 nm to determine skeletal muscle oxygen saturation. Previous studies have used the ratio of both signals to report muscle oxygen saturation.

Purpose: The purpose of this pilot study is to assess the different approaches used to represent muscle oxygen saturation and to evaluate the pulsations of oxygenated hemoglobin/myoglobin (Oheme) and deoxygenated hemoglobin/myoglobin (Heme) signals.

Method: Twelve participants, aged 20-29 years, were tested on the forearm flexor muscles using continuous-wave NIRS at rest. Measurements were taken during 2-3 min rest, physiological calibration (5 min ischemia), and reperfusion. Ten participants were included in the study analysis.

Results: There was a significant difference in pulse size between Oheme and Heme signals at the three locations ( < 0.05). Resting oxygen saturation was 58.8% + 9.2%, 69.6% + 3.9%, and 89.2% + 6.9% when calibrated using Oheme, the tissue oxygenation/saturation index (TSI), and Heme, respectively.

Conclusion: The difference in magnitude of Oheme and Heme pulses with each heartbeat might suggest different anatomical locations of these signals, for which calibrating with just one of the signals instead of the ratio of both is proposed. Calculations of physiological calibration must account for increased blood volume in the tissue because of the changes in blood volume, which appear to be primarily from the Oheme signal. Resting oxygen levels calibrated with Heme agree with theoretical oxygen saturation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7739319PMC
http://dx.doi.org/10.3390/jfmk4020028DOI Listing

Publication Analysis

Top Keywords

oxygen saturation
16
muscle oxygen
12
near-infrared spectroscopy
8
spectroscopy nirs
8
skeletal muscle
8
physiological calibration
8
oheme heme
8
resting oxygen
8
blood volume
8
oxygen
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!