Uniform dose distribution with steep lateral gradient within depth range of 0-0.5 cm is crucial to be able to treat small skin lesions. The standard nominal 4 MeV electron beam from Elekta Versa HD linear accelerator was modified with degrading filter to remove the lateral scatter from treatment head and minimize the penumbra. The energy degrading method was verified based on dosimetric properties and output factors (OFs) with comparison of four types of measurement methods. The properties of degraded 4 MeV electron beam and developed electron applicators seem optimal for treating small targets near the skin surface.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807873PMC
http://dx.doi.org/10.1016/j.phro.2019.04.003DOI Listing

Publication Analysis

Top Keywords

4 mev electron
12
electron beam
12
linear accelerator
8
modification 4 mev
4
electron
4
beam linear
4
accelerator irradiation
4
irradiation small
4
small superficial
4
superficial skin
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!