Background: In the emergency department (ED) setting, rapid testing for SARS-CoV-2 is likely associated with advantages to patients and healthcare workers, for example, enabling early but rationale use of limited isolation resources. Most recently, several SARS-CoV-2 rapid point-of-care antigen tests (AGTEST) became available. There is a growing need for data regarding their clinical utility and performance in the diagnosis of SARS-CoV-2 infection in the real life setting EDs.

Methods: We implemented AGTEST (here: Roche/SD Biosensor) in all four adult and the one paediatric EDs at Charité - Universitätsmedizin Berlin in our diagnostic testing strategy. Test indication was limited to symptomatic suspected COVID-19 patients. Detailed written instructions on who to test were distributed and testing personnel were trained in proper specimen collection and handling. In each suspected COVID-19 patient, two sequential deep oro-nasopharyngeal swabs were obtained for viral tests. The first swab was collected for nucleic acid testing through SARS-CoV-2 real-time reverse transcriptase (rt)-PCR diagnostic panel (PCRTEST) in the central laboratory. The second swab was collected to perform the AGTEST. Analysis of routine data was prospectively planned and data were retrieved from the medical records after the inclusion period in the adult or paediatric ED. Diagnostic performance was calculated using the PCRTEST as reference standard. False negative and false positive AGTEST results were analysed individually and compared with viral concentrations derived from the calibrated PCRTEST.

Results: We included  = 483 patients including  = 202 from the paediatric ED.  = 10 patients had to be excluded due to missing data and finally  = 473 patients were analysed. In the adult cohort, the sensitivity of the AGTEST was 75.3 (95%CI: 65.8/83.4)% and the specificity was 100 (95%CI: 98.4/100)% with a SARS-CoV-2 prevalence of 32.8%; the positive predictive value was 100 (95%CI: 95.7/100)% and the negative predictive value 89.2 (95%CI: 84.5/93.9)%. In the paediatric cohort, the sensitivity was 72.0 (95%CI: 53.3/86.7)%, the specificity was 99.4 (95%CI:97.3/99.9)% with a prevalence of 12.4%; the positive predictive value was 94.7 (95%CI: 78.3/99.7)% and the negative predictive value was 96.2 (95%CI:92.7/98.3)%. Thus,  = 22 adult and  = 7 paediatric patients showed false negative AGTEST results and only one false positive AGTEST occurred, in the paediatric cohort. Calculated viral concentrations from the rt-PCR lay between 3.16 and 9.51 log10 RNA copies/mL buffer. All false negative patients in the adult ED cohort, who had confirmed symptom onset at least seven days earlier had less than 5 × 10 RNA copies/mL buffer.

Conclusions: We conclude that the use of AGTEST among symptomatic patients in the emergency setting is useful for the early identification of COVID-19, but patients who test negative require confirmation by PCRTEST and must stay isolated until this result becomes available. Adult patients with a false negative AGTEST and symptom onset at least one week earlier have typically a low SARS-CoV-2 RNA concentration and are likely no longer infectious.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7898296PMC
http://dx.doi.org/10.1080/1354750X.2021.1876769DOI Listing

Publication Analysis

Top Keywords

false negative
16
patients
10
agtest
9
emergency department
8
testing sars-cov-2
8
adult paediatric
8
suspected covid-19
8
covid-19 patients
8
swab collected
8
false positive
8

Similar Publications

cutaneous melanoma has often unpredictable lymphatic drainage patterns, especially at the level of the trunk, head and neck regions. Sentinel lymph node biopsy (SLNB) is an important prognostic tool that accurately assesses regional lymph node involvement and guides therapeutic decisions. Material and this prospective study involved 104 patients diagnosed with cutaneous melanoma who underwent SLNB using a radioactive tracer.

View Article and Find Full Text PDF

Background: Anxiety disorders are common in alcohol use disorder (AUD) treatment patients. Such co-occurring conditions ("comorbidity") have negative prognostic implications for AUD treatment outcomes, yet they commonly go unaddressed in standard AUD care. Over a decade ago, we developed and validated a cognitive behavioral therapy intervention to supplement standard AUD care that, when delivered by trained therapists, improves outcomes in comorbid patients.

View Article and Find Full Text PDF

Background: Early posttransplant cytomegalovirus (CMV) infections in CMV seronegative solid organ transplant recipients (SOTR) with CMV seronegative donors (D-/R-) are often attributed transfusion-transmitted CMV. The prevalence of false-negative donor CMV serology in D-/R- SOTR with early CMV infections has not been explored.

Methods: We determined the frequency and characteristics of CMV DNAemia that occurred within 90 days of transplant among adult SOTR classified as D-/R- who underwent a first SOT at a single center between February 25, 2014 and February 25, 2024.

View Article and Find Full Text PDF

Photoelectrochemical biosensing platform based on switch of photocurrent polarity of CdS/Ni-CAT nanorod arrays by Au@CuO for highly selective and sensitive CEA detection.

Talanta

December 2024

State Key Laboratory for Chemistry and Molecular Engineering of Medicinal Resources, Key Laboratory for Chemistry and Molecular Engineering of Medicinal Resources (Ministry of Education of China), Collaborative Innovation Center for Guangxi Ethnic Medicine, School of Chemistry and Pharmaceutical Sciences, Guangxi Normal University, Guilin, 541004, PR China. Electronic address:

Sensitive and accurate determination of tumour biomarkers is extremely important for early cancer diagnosis. Herein, a photoelectrochemical biosensor platform was constructed for ultrasensitive tumour biomarker detection by utilizing Au@CuO to switch the photocurrent polarity of CdS/Ni-catecholates metal-organic framework (Ni-CAT) nanorod arrays grown in situ on ITO. The Ni-CAT obtains close contact with ITO and forms a Z-scheme heterojunction with CdS, which improves the photogenerated electron transfer ability.

View Article and Find Full Text PDF

Electrochemical Biosensors for Cancer Diagnosis: Multitarget Analysis to Present Molecular Characteristics of Tumor Heterogeneity.

JACS Au

December 2024

Center for Molecular Recognition and Biosensing, Shanghai Engineering Research Center of Organ Repair, Joint International Research Laboratory of Biomaterials and Biotechnology in Organ Repair (Ministry of Education), School of Life Sciences, Shanghai University, Shanghai 200444, China.

Electrochemical biosensors are gaining attention as powerful tools in cancer diagnosis, particularly in liquid biopsy, due to their high efficiency, rapid response, exceptional sensitivity, and specificity. However, the complexity of intra- and intertumor heterogeneity, with variations in genetic and protein expression profiles and epigenetic modifications, makes electrochemical biosensors susceptible to false-positive or false-negative diagnostic outcomes. To address this challenge, there is growing interest in simultaneously analyzing multiple biomarkers to reveal molecular characteristics of tumor heterogeneity for precise cancer diagnosis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!