  • * It categorizes the femurs into four groups based on nail diameter, fracture type, and distal filling, and examines nail movement and erosion in specific femoral areas under cyclic loading.
  • * Results show that smaller nail diameters and less stable fractures result in more nail movement, which in turn increases the risk of varus collapse, highlighting the need for solid nail fitting to minimize movement.

Article Abstract

Objective: Cephalomedullary nails are tools commonly used for surgical fixation in proximal femoral fractures. The most common complications in their use are varus collapse and screw cutout of the femoral head. The objective of this study is to examine the effects of intramedullary nail movement on varus collapse and screw cutout.

Materials And Methods: The study was conducted on 70 composite femur models treated with PFNA. We divided the femurs into 4 groups based on the differences in nail diameter, fracture type and filling of the distal intramedullary area. All femurs were exposed to axial cyclic loading. Each femur was examined in terms of intramedullary nail movement and amount of erosions in femoral medulla [amount of erosion in femoral head (FT distance), amount of erosion in femoral neck (FB distance), amount of erosion in trochanteric major (TB distance) and expansion of trochanter tip (TT distance)].

Results: We found that degree of nail movement in the intramedullary region was inversely correlated with nail diameter and directly correlated with instability of fracture. One of the parameters used to evaluate varus development, FB distance, was affected by the degree of intramedullary nail movement and fracture type. TB distance was affected by nail diameter.

Conclusions: Nail diameter and fracture type are effective in intramedullary nail movement. Varus collapse progress is accelerated by the increase in nail movement in the intramedullary region. Thus, we conclude that it is important to strengthen diaphyseal adherence, which decreases intramedullary movement of the nail.

Level Of Evidence: III; well-design case control study.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.otsr.2021.102804DOI Listing

Publication Analysis

Top Keywords

nail movement
24
varus collapse
16
intramedullary nail
16
nail diameter
12
fracture type
12
nail
10
cephalomedullary nails
8
collapse screw
8
femoral head
8
intramedullary
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!