Cardiac I-MIBG planar heart to mediastinum ratios depend on patient size; phantom studies suggest SPECT-CT could improve quantification.

Biomed Phys Eng Express

Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle Upon Tyne. NE4 6BE, United Kingdom. Nuclear Medicine Department, Leazes Wing, Royal Victoria Infirmary, Richardson Road, Newcastle Upon Tyne. NE1 4LP, United Kingdom.

Published: December 2019

Introduction: Planar I-MIBG (meta-iodobenzylguanidine) cardiac imaging is listed as an indicative biomarker in the 2017 international consensus criteria for the diagnosis of dementia with Lewy bodies. There has been very little research into the relationship between apparent cardiac uptake and patient size, or in the possible advantage of attenuation and scatter corrected SPECT-CT compared to planar imaging. We aimed to evaluate this in both a chest phantom and in older adults with normal cognition.

Materials And Methods: An anthropomorphic chest phantom was filled with I solution using activities typical of healthy subjects. The phantom was scanned on a Siemens Intevo gamma camera with MELP collimators using both planar and SPECT-CT techniques. Further scans were acquired with a PMMA chest plate added, then water filled plastic breasts. The SPECT-CT images were reconstructed using a resolution recovery OSEM method with and without attenuation and scatter correction (ACSC) applied. Twenty-nine adults over 60 years of age (mean 75.2 ± 8.3 years) underwent planar cardiac MIBG imaging, followed by SPECT-CT. SPECT images were reconstructed as above. Heart-to-mediastinum ratios (HMRs) were calculated for planar and SPECT images.

Results: Phantom planar HMR decreased by 20% with the PMMA chest plate added; 39% with plate and breasts. ACSC SPECT cardiac counts showed less dependence on phantom size than SPECT without ACSC (3% versus 37%). The body mass indices (BMI) of the older adults ranged from 22 to 38. There was a significant linear relationship between planar HMR and BMI (R = 0.44, p<0.01), but not for ACSC SPECT. However, there was no significant difference between the slopes for planar and ACSC SPECT (p = 0.11).

Conclusion: Planar cardiac I-MIBG HMR results are correlated with BMI. Phantom results suggest that ACSC SPECT can correct for patient size. A large patient population or clinical database would be required to demonstrate a clinical effect.

Download full-text PDF

Source
http://dx.doi.org/10.1088/2057-1976/ab5c09DOI Listing

Publication Analysis

Top Keywords

planar
8
patient size
8
attenuation scatter
8
chest phantom
8
older adults
8
pmma chest
8
chest plate
8
images reconstructed
8
planar hmr
8
phantom
6

Similar Publications

Dual-polarity readout is a simple and robust way to mitigate Nyquist ghosting in diffusion-weighted echo-planar imaging but imposes doubled scan time. We here propose how dual-polarity readout can be implemented with little or no increase in scan time by exploiting an observed b-value dependence and signal averaging. The b-value dependence was confirmed in healthy volunteers with distinct ghosting at low b-values but of negligible magnitude at b = 1000 s/mm2.

View Article and Find Full Text PDF

Background: Central arterial stiffening is associated with brain white matter (WM) damage and gray matter (GM) volume loss in older adults, but little is known about this association from an adult lifespan perspective.

Purpose: To investigate the associations of central arterial stiffness with WM microstructural organization, WM lesion load, cortical thickness, and GM volume in healthy adults across the lifespan.

Study Type: This is a cross-sectional study.

View Article and Find Full Text PDF

Patterning Planar, Flexible Li-S Battery Full Cells on Laser-Induced Graphene Traces.

Nanomaterials (Basel)

December 2024

Quantum Nano Centre, Department of Chemical Engineering, University of Waterloo, Waterloo, ON N2L 3G1, Canada.

Laser conversion of commercial polymers to laser-induced graphene (LIG) using inexpensive and accessible CO lasers has enabled the rapid prototyping of promising electronic and electrochemical devices. Frequently used to pattern interdigitated supercapacitors, few approaches have been developed to pattern batteries-in particular, full cells. Herein, we report an LIG-based approach to a planar, interdigitated Li-S battery.

View Article and Find Full Text PDF

In neurons, the acquisition of a polarized morphology is achieved upon the outgrowth of a single axon from one of several neurites. Small extracellular vesicles (sEVs), such as exosomes, from diverse sources are known to promote neurite outgrowth and thus may have therapeutic potential. However, the effect of fibroblast-derived exosomes on axon elongation in neurons of the central nervous system under growth-permissive conditions remains unclear.

View Article and Find Full Text PDF

Significant progress has been made through the optimization of modelling and device architecture solar cells has proven to be a valuable and highly effective approach for gaining a deeper understanding of the underlying physical processes in solar cells. Consequently, this research has conducted a two-dimensional (2D) perovskite solar cells (PSCs) simulation to develop an accurate model. The approach utilized in this study is based on the finite element method (FEM).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!