Purpose: To prospectively investigate the impact of image reconstruction on MRI radiomics features.
Methods: An anthropomorphic phantom was scanned at 1.5 T using a standardized sequence for MR-guided radiotherapy under SENSE and compressed-SENSE reconstruction settings. A total of 93 first-order and texture radiomics features in 10 volumes of interest were assessed based on (1) accuracy measured by the percentage deviation from the reference, (2) robustness on reconstruction in all volumes of interest measured by the intraclass correlation coefficient, and (3) repeatability measured by the coefficient of variance over the repetitive acquisitions. Finally, reliable and unreliable radiomics features were comprehensively determined based on their accuracy, robustness, and repeatability.
Results: Better accuracy and robustness of the radiomics features were achieved under SENSE than compressed-SENSE reconstruction. The feature accuracy under SENSE reconstruction was more affected by acceleration factor than direction, whereas under compressed-SENSE reconstruction, accuracy was substantially impacted by the increasing denoising levels. Feature repeatability was dependent more on feature types than on reconstruction. A total of 45 reliable features and 13 unreliable features were finally determined for SENSE, compared with 22 reliable and 26 unreliable features for compressed SENSE. First-order and gray-level co-occurrence matrix features were generally more reliable than other features.
Conclusion: Radiomics features could be substantially affected by MRI reconstruction, so precautions need to be taken regarding their reliability for clinical use. This study helps the guidance of the preselection of reliable radiomics features and the preclusion of unreliable features in MR-guided radiotherapy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mrm.28650 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!