Objective: To demonstrate the performance and safety of a bihormonal (insulin and glucagon) artificial pancreas (AP) in adults with type 1 diabetes.

Research Design And Methods: In this outpatient, randomized, crossover trial, 2-week fully closed loop glucose control (AP therapy) was compared with 2-week open loop control (patient's normal insulin pump therapy with a glucose sensor if they had one).

Results: A total of 23 patients were included in the analysis. Time in range (70-180 mg/dL [3.9-10 mmol/L]) was significantly higher during closed loop (median 86.6% of time [interquartile range 84.9-88.5]) compared with open loop (53.9% [49.7-67.2]; < 0.0001).

Conclusions: Compared with insulin pump therapy, the bihormonal AP provided superior glucose control, without meal or exercise announcements, and was safe in adults with type 1 diabetes.