: This study aimed to evaluate the efficacy and safety of a newly developed, leak closure Internal Compression Therapy (ICT) (Invamed, Ankara, Turkey) device during a single-session procedure in a group of patients with primary deep valve incompetence. : There were 286 patients who were diagnosed with deep venous reflux by duplex scanning. They underwent valvular leak operations to treat primary deep venous insufficiency. Follow-up visits were on the third day, first month, sixth month and twelfth month. At each visit, duplex scanning and a clinical examination were performed. Successful treatment was defined as deep vein valves without reflux. Any patency or reflux over 1 sec was considered a failure. : The study enrolled 286 patients with deep venous insufficiency. Procedural technical success was 100%. At the one-year follow-up, the overall success, among all patients, was 92%. No significant morbidity or mortality related to the procedure were observed. All patients had major improvements in venous clinical severity score (VCSS) scores postoperatively. VCSS scores at pre-intervention, and at the twelfth month, were 20.7±5.9 and 3.9±0.9, respectively (p<0.001). : After the twelve-month follow-up, the postprocedural outcomes indicate the ICT device is safe and effective.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7751084PMC
http://dx.doi.org/10.3400/avd.oa.20-00016DOI Listing

Publication Analysis

Top Keywords

deep venous
12
deep vein
8
primary deep
8
286 patients
8
duplex scanning
8
venous insufficiency
8
twelfth month
8
vcss scores
8
deep
6
patients
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!