Background: The supplemental screw technique was introduced for salvage of lateral hinge fracture in medial open-wedge high tibial osteotomy (owHTO). The efficacy of its use in protection of lateral hinge fracture and corresponding biomechanical behaviors remained unclear. The current study was aimed to clarify if a supplemental screw can provide better protection to lateral hinge in biomechanical perspective.

Materials: An in vitro biomechanical test was conducted. Tibial sawbones, commercial owHTO plates and a cannulated screw were utilized for preparing the intact, owHTO, and owHTO with cannulated screw insertion specimens. A "staircase" dynamic load protocol was adopted for axial compressive test with increasing load levels to determine structural strength and durability by using a material testing system, while a motion capture system was applied for determining the dynamic changes in varus angle and posterior slope of the tibia plateau with various specimen preparation conditions.

Results: Type II lateral hinge fracture were the major failure pattern in all specimens prepared with owHTO. The insertion of a supplemental cannulated screw in medial owHTO specimens reinforced structural stability and durability in dynamic cyclic loading tests: the compressive stiffness increased to 58.9-62.2% of an intact specimen, whereas the owHTO specimens provided only 23.7-29.2% of stiffness of an intact specimen. In view of tibial plateau alignment, the insertion of a supplemental screw improved the structural deficiency caused by owHTO, and reduced the posterior slope increase and excessive varus deformity by 81.8% and 83.2%, respectively.

Conclusion: The current study revealed that supplemental screw insertion is a simple and effective technique to improve the structural stability and durability in medial owHTO.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7773260PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0244557PLOS

Publication Analysis

Top Keywords

supplemental screw
20
lateral hinge
16
hinge fracture
12
cannulated screw
12
owhto
9
medial open-wedge
8
open-wedge high
8
high tibial
8
tibial osteotomy
8
protection lateral
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!