This study aimed to evaluate the influence of large apical preparations with Reciproc (REC), Hyflex CM (HCM) and Twisted File Adaptive (TFA) systems using micro-computed tomography (MCT). Ninety mesiobuccal (MB) and distobuccal (DB) root canals of maxillary molars (n = 45) were scanned using MCT before and after the shaping procedures. The root canals (n = 15) were prepared until REC 40.06, HCM 40.04 and TFA 35.04. The root canal transportation (RCT), centring ability (CA), change in volume of the root canal and at different levels (VC), remaining dentine thickness (RDT), removal of dentine wall (RDW) and working time (WT) were evaluated. Data were analysed using the Kruskal-Wallis and Dunn tests, and the one-way ANOVA and Tukey tests with a level of significance set at 5%. No significant difference among the instruments was found regarding the RCT, CA, RDT, RDW and WT (P > 0.05), in larger apical preparations in curved MB and straight DB canals of maxillary molars.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aej.12481DOI Listing

Publication Analysis

Top Keywords

canals maxillary
12
maxillary molars
12
apical preparations
8
root canals
8
root canal
8
safety large
4
large preparation
4
preparation instruments
4
instruments buccal
4
canals
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!