SpheriCal -ESI: A dendrimer-based nine-point calibration solution ranging from m/z 273 to 1716 for electrospray ionization mass spectrometry peptide analysis.

Rapid Commun Mass Spectrom

School of Engineering Sciences in Chemistry, Biotechnology and Health, Department of Chemistry, Division of Applied Physical Chemistry, Analytical Chemistry, KTH Royal Institute of Technology, Stockholm, Sweden.

Published: March 2021

Rationale: A calibration solution for mass spectrometry needs to cover the range of interest with intense and sufficiently narrowly spaced peaks. Limited options fulfilling this may lead to compromises between performance and ease of use. SpheriCal -ESI was designed to combine high calibration performance for electrospray ionization (ESI) mass spectrometric analysis of peptides in positive mode with quick and easy use.

Methods: The developed calibration solution was tested using three mass spectrometers: two ion traps and one tandem quadrupole. The m/z errors of SpheriCal -ESI itself and of a tryptic digest of cytochrome C were measured after calibration. The results were compared with those achieved with ESI Tuning Mix. The memory effects of the dendrimers, and contamination from Na in the calibration solution, were evaluated.

Results: SpheriCal -ESI showed good shelf life as powder and was quickly reconstituted for use. Achieving intense and stable signals was straightforward. The accuracies and precisions were as expected for the instruments. SpheriCal -ESI was more precise and at least as accurate as ESI Tuning Mix. The memory effects and Na contamination were found to be negligible in typical peptide solvents. In addition, the dendrimers showed predictable dissociations with product ions common to collision-induced dissociation in both ion trap and tandem quadrupole mass spectrometers.

Conclusions: SpheriCal -ESI provided easily accessible calibration by showing intense signals at low infusion rates and at source settings equal or similar to those used in peptide analysis. Nine calibration points in the range of interest gave precise and accurate results. Memory effects and contamination were negligible even without rinsing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7900979PMC
http://dx.doi.org/10.1002/rcm.9035DOI Listing

Publication Analysis

Top Keywords

spherical -esi
24
calibration solution
16
memory effects
12
calibration
8
electrospray ionization
8
mass spectrometry
8
peptide analysis
8
range interest
8
tandem quadrupole
8
esi tuning
8

Similar Publications

, has been widely recognized for its medical applications. LC-ESI-TOF-MS identified 22 secondary metabolites including phenolics, flavonoids, and anthocyanidin glycosides among its total extract (LCTE). The study aimed to apply LCTE as a biogenic material for reducing and capping the silver nanoparticles (LC-AgNPs).

View Article and Find Full Text PDF

The increasing emergence of multidrug-resistant (MDR) pathogens due to antibiotic misuse translates into obstinate infections with high morbidity and high-cost hospitalizations. To oppose these MDR superbugs, new antimicrobial options are necessary. Although both quaternary ammonium salts (QASs) and phosphonium salts (QPSs) possess antimicrobial effects, QPSs have been studied to a lesser extent.

View Article and Find Full Text PDF

α-Ketoglutaric acid-based supramolecular Zn(II) metallogels in ,'-dimethylformamide (DMF) and dimethyl sulfoxide (DMSO) solvent (i.e., Zn-α-Glu-DMF and Zn-α-Glu-DMSO) were successfully achieved.

View Article and Find Full Text PDF

Worldwide meat consumption and production have nearly quintupled in the last 60 years. In this context, research and the application of new technologies related to animal reproduction have evolved in an accelerated way. The objective of the present study was to apply nanoemulsions (NEs) as carriers of lipids to feed bovine embryos in culture media and verify their impact on the development of embryos produced .

View Article and Find Full Text PDF

This study uses real-time monitoring, at microsecond time scales, with a charge-sensing particle detector to investigate the evaporation and fission processes of methanol/micrometer-sized polystyrene beads (PS beads) droplets and bacterial particles droplets generated via electrospray ionization (ESI) under elevated temperatures. By incrementally raising capillary temperatures, the solvent, such as methanol on 0.75 μm PS beads, experiences partial evaporation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!