Purpose: In multileaf collimator (MLC) tracking, the MLC positions from the original treatment plan are continuously modified to account for intrafraction tumor motion. As the treatment is adapted in real time, there is additional risk of delivery errors which cannot be detected using traditional pretreatment dose verification. The purpose of this work is to develop a system for real-time geometric verification of MLC tracking treatments using an electronic portal imaging device (EPID).

Methods: MLC tracking was utilized during volumetric modulated arc therapy (VMAT). During these deliveries, treatment beam images were taken at 9.57 frames per second using an EPID and frame grabber computer. MLC positions were extracted from each image frame and used to assess delivery accuracy using three geometric measures: the location, size, and shape of the radiation field. The EPID-measured field location was compared to the tumor motion measured by implanted electromagnetic markers. The size and shape of the beam were compared to the size and shape from the original treatment plan, respectively. This technique was validated by simulating errors in phantom test deliveries and by comparison between EPID measurements and treatment log files. The method was applied offline to images acquired during the LIGHT Stereotactic Ablative Body Radiotherapy (SABR) clinical trial, where MLC tracking was performed for 17 lung cancer patients. The EPID-based verification results were subsequently compared to post-treatment dose reconstruction.

Results: Simulated field location errors were detected during phantom validation tests with an uncertainty of 0.28 mm (parallel to MLC motion) and 0.38 mm (perpendicular), expressed as a root-mean-square error (RMS ). For simulated field size errors, the RMS was 0.47 cm and field shape changes were detected for random errors with standard deviation ≥ 2.5 mm. For clinical lung SABR deliveries, field location errors of 1.6 mm (parallel MLC motion) and 4.9 mm (perpendicular) were measured (expressed as a full-width-half-maximum). The mean and standard deviation of the errors in field size and shape were 0.0 ± 0.3 cm and 0.3 ± 0.1 (expressed as a translation-invariant normalized RMS). No correlation was observed between geometric errors during each treatment fraction and dosimetric errors in the reconstructed dose to the target volume for this cohort of patients.

Conclusion: A system for real-time delivery verification has been developed for MLC tracking using time-resolved EPID imaging. The technique has been tested offline in phantom-based deliveries and clinical patient deliveries and was used to independently verify the geometric accuracy of the MLC during MLC tracking radiotherapy.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14675DOI Listing

Publication Analysis

Top Keywords

mlc tracking
28
size shape
16
mlc
12
field location
12
errors
9
verification mlc
8
tracking treatments
8
time-resolved epid
8
epid imaging
8
mlc positions
8

Similar Publications

Development and validation of the Normalized Organoid Growth Rate (NOGR) metric in brightfield imaging-based assays.

Commun Biol

December 2024

Center for Oncological Research (CORE), Integrated Personalized & Precision Oncology Network (IPPON), University of Antwerp, Wilrijk, Belgium.

This study focuses on refining growth-rate-based drug response metrics for patient-derived tumor organoid screening using brightfield live-cell imaging. Traditional metrics like Normalized Growth Rate Inhibition (GR) and Normalized Drug Response (NDR) have been used to assess organoid responses to anticancer treatments but face limitations in accurately quantifying cytostatic and cytotoxic effects across varying growth rates. Here, we introduce the Normalized Organoid Growth Rate (NOGR) metric, specifically developed for brightfield imaging-based assays.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the effectiveness of using 3D-spheroids of human adipose stem cells (hASCs) for treating fecal incontinence by focusing on their ability to develop into smooth muscle cells (SMCs) when implanted in a mouse model.
  • Researchers created a targeted model by freezing part of the internal anal sphincter (IAS) in rats and then analyzed SMC markers and related proteins after implanting either 3D-spheroids or 2D-cultured cells.
  • Results showed that 3D-spheroids had greater expression of SMC markers and induced better contractile characteristics over time compared to 2D-hASCs, indicating a potential therapeutic benefit for fecal incontinence treatment.
View Article and Find Full Text PDF

Background: Canada has one of the highest incidences of colorectal cancer (CRC) worldwide. CRC screening improves CRC outcomes and is cost-effective. This study compares Canadian CRC screening programs using essential elements of an organized screening program outlined by the International Agency for Research on Cancer (IARC).

View Article and Find Full Text PDF

Purpose: This study presents a novel and comprehensive framework for evaluating magnetic resonance guided radiotherapy (MRgRT) workflow by integrating the Failure Modes and Effects Analysis (FMEA) approach with Time-Driven Activity-Based Costing (TDABC). We assess the workflow for safety, quality, and economic implications, providing a holistic understanding of the MRgRT implementation. The aim is to offer valuable insights to healthcare practitioners and administrators, facilitating informed decision-making regarding the 0.

View Article and Find Full Text PDF

Background: Several (online) adaptive radiotherapy procedures are available to maximize healthy tissue sparing in the presence of inter/intrafractional motion during stereotactic body radiotherapy (SBRT) on an MR-linac. The increased treatment complexity and the motion-delivery interplay during these treatments require MR-compatible motion phantoms with time-resolved dosimeters to validate end-to-end workflows. This is not possible with currently available phantoms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!