Background: It is recommended to have 6 bicortical screws for plate fixation of long bone fractures; however, many metacarpal fractures do not allow 6 screws due to size limitations and proximity of crucial anatomical structures. The purpose of this biomechanical study was to determine whether the mechanical properties of a 4-screw nonlocking construct are noninferior to those of a 6-screw nonlocking construct.
Methods: Metacarpal sawbones were used to simulate a midshaft, transverse fracture. Nonlocking bicortical screws were placed in the 6-hole plate, and the metacarpals were randomly assigned to 2 equal study groups: (1) 4 screws, 2 on either side of the fracture (4S); and (2) 6 screws, 3 on either side of the fracture (6S). The metacarpals were tested in a cyclic loading mode and load to failure in a cantilever bending mode.
Results: Maximum deflection was significantly higher for 4S compared with 6S. Cyclic root mean square (RMS) was also significantly greater for 4S at 70 and 100 N. There were no statistically significant differences observed between the 2 constructs for maximum bending load, bending stiffness, and cyclic RMS at 40 N. The maximum bending load in 4S and 6S was 245.6 ± 37.9 N and 230.8 ± 41.9 N, respectively; 4S was noninferior and not superior to 6S. Noninferiority testing was inconclusive for bending stiffness.
Conclusions: A 4-screw bicortical nonlocking construct is noninferior to a 6-screw bicortical nonlocking construct for fixation of metacarpal fractures, which may be advantageous to minimize disruption of soft tissues while maintaining sufficient construct stability.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9465802 | PMC |
http://dx.doi.org/10.1177/1558944720974116 | DOI Listing |
Acta Orthop
November 2024
Department of Trauma, Hand and Reconstructive Surgery, University Hospital Münster, Münster, Germany.
J Clin Orthop Trauma
June 2024
University of Alabama Birmingham Department of Orthopaedics, Birmingham, AL, United States.
Introduction: Distal fibula osteoporotic comminuted fractures are challenging to treat and are often treated with periarticular locking plates. This study examined the biomechanical difference between locked plating and dual non-locked one-third tubular plating.
Methods: Using an osteoporotic Sawbones fibula model, simulated fracture were fixated with one-third tubular dual plating and locked periarticular plating.
Arthrosc Sports Med Rehabil
June 2024
MedStar Orthopaedic Institute, Union Memorial Hospital, Baltimore, Maryland, U.S.A.
Purpose: To evaluate the biomechanical properties of open versus percutaneous Achilles tendon repair.
Methods: A systematic review of original research articles was performed using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. To qualify for study inclusion, articles were required to be published in English, use a laboratory design using either human or animal tissue, and directly compare the biomechanical properties of open Achilles repair using a Krackow or Kessler technique with percutaneous repair using either a locking or nonlocking suture construct.
J Clin Med
May 2024
Biomechanics Laboratory, Legacy Research Institute, Portland, OR 97232, USA.
The controlled dynamization of fractures can promote natural fracture healing by callus formation, while overly rigid fixation can suppress healing. The advent of locked plating technology enabled new strategies for the controlled dynamization of fractures, such as far cortical locking (FCL) screws or active plates with elastically suspended screw holes. However, these strategies did not allow for the use of non-locking screws, which are typically used to reduce bone fragments to the plate.
View Article and Find Full Text PDFOTA Int
June 2024
Orthopaedic Surgery Department, Boston University Medical Center, Boston, MA.
Objectives: The objective of this study was to compare the biomechanical properties of locked and nonlocked diaphyseal fixation in a model of distal femur fractures using osteoporotic and nonosteoporotic human cadaveric bone.
Methods: A supracondylar osteotomy was created to mimic a fracture (OTA/AO 33A3) in osteoporotic (n = 4) and nonosteoporotic (n = 5) cadaveric distal femurs. The left and right femurs of each pair were instrumented with a distal femoral locking plate and randomly assigned to have diaphyseal fixation with either locked or nonlocked screws.