Methods for the determination of CS-570, a chemically stable prostacyclin analogue, in plasma were developed by using an immobilized antibody column followed by fluorescence HPLC and GC/MS. The CS-570 antibody, obtained from rabbit plasma by giving CS-570-BSA for a few months, was coupled to Sepharose 4B and used as extraction phase for sample clean-up and extraction of the drug. A plasma sample was applied to this column, washed with water and the drug was eluted with 90% acetonitrile. 0.02% (w/v) 9-anthryldiazomethane (ADAM) was added to the extract to form a fluorescent derivative. The CS-570-ADAM adduct exhibited high sensitivity when applied to HPLC with fluorescence detection and column switching. The detection limit was 1 ng/mL when 1 mL of plasma was available. Additionally, a pentafluorobenzyltrimethylsilyl derivative of CS-570 showed excellent sensitivity when determined by capillary GC interfaced to negative ion chemical ionization MS using a stable isotope labelled analogue as an internal standard.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/bmc.1130020507 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!