Background: We aimed to evaluate whether the use of our novel patient-specific guide (PSG) with 3-dimensional reconstruction in reverse total shoulder arthroplasty (RTSA) would allow accurate and reliable implantation of the glenoid and humeral components.

Methods: 20 fresh-frozen cadaveric shoulders were used. The PSG group (n=10) and conventional group (n=10) was evaluated the accuracy and reproducibility of implant positioning between before and after surgery on the computed tomography image.

Results: The superoinferior and anteroposterior offset in the glenoid component were 0.42 ± 0.07, 0.50 ± 0.08 in the conventional group and 0.45 ± 0.03, 0.46 ± 0.02 in the PSG group. The inclination and version angles were -1.93° ± 4.31°, 2.27° ± 5.91° and 0.46° ± 0.02°, 3.38° ± 2.79°. The standard deviation showed a smaller difference in the PSG group. The anteroposterior and lateromedial humeral canal center offset in the humeral component were 0.45 ± 0.12, 0.48 ± 0.15 in the conventional group and 0.46 ± 0.59 (=0.794), 0.46 ± 0.06 (=0.702) in the PSG group. The PSG showed significantly better humeral stem alignment.

Conclusions: The use of PSGs with 3-dimensional reconstruction reduces variabilities in glenoid and humerus component positions and prevents extreme positioning errors in RTSA.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7713874PMC
http://dx.doi.org/10.5397/cise.2019.22.1.16DOI Listing

Publication Analysis

Top Keywords

psg group
16
3-dimensional reconstruction
12
conventional group
12
accuracy reproducibility
8
reverse total
8
total shoulder
8
shoulder arthroplasty
8
group n=10
8
group
7
psg
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!