Aim: Attenuation correction using zero-echo time (ZTE) - magnetic resonance imaging (MRI) (ZTE-MRAC) has become one of the standard methods for brain-positron emission tomography (PET) on commercial PET/MR scanners. Although the accuracy of the net tracer-uptake quantification based on ZTE-MRAC has been validated, that of the diagnosis for dementia has not yet been clarified, especially in terms of automated statistical analysis. The aim of this study was to clarify the impact of ZTE-MRAC on the diagnosis of Alzheimer's disease (AD) by performing simulation study.

Methods: We recruited 27 subjects, who underwent both PET/computed tomography (CT) and PET/MR (GE SIGNA) examinations. Additionally, we extracted 107 subjects from the Alzheimer Disease Neuroimaging Initiative (ADNI) dataset. From the PET raw data acquired on PET/MR, three FDG-PET series were generated, using two vendor-provided MRAC methods (ZTE and Atlas) and CT-based AC. Following spatial normalization to Montreal Neurological Institute (MNI) space, we calculated each patient's specific error maps, which correspond to the difference between the PET image corrected using the CTAC method and the PET images corrected using the MRAC methods. To simulate PET maps as if ADNI data had been corrected using MRAC methods, we multiplied each of these 27 error maps with each of the 107 ADNI cases in MNI space. To evaluate the probability of AD in each resulting image, we calculated a cumulative -value using a fully automated method which had been validated not only in the original ADNI dataset but several multi-center studies. In the method, PET score = 1 is the 95% prediction limit of AD. PET score and diagnostic accuracy for the discrimination of AD were evaluated in simulated images using the original ADNI dataset as reference.

Results: Positron emission tomography score was slightly underestimated both in ZTE and Atlas group compared with reference CTAC (-0.0796 ± 0.0938 vs. -0.0784 ± 0.1724). The absolute error of PET score was lower in ZTE than Atlas group (0.098 ± 0.075 vs. 0.145 ± 0.122, < 0.001). A higher correlation to the original PET score was observed in ZTE vs. Atlas group ( : 0.982 vs. 0.961). The accuracy for the discrimination of AD patients from normal control was maintained in ZTE and Atlas compared to CTAC (ZTE vs. Atlas. vs. original; 82.5% vs. 82.1% vs. 83.2% (CI 81.8-84.5%), respectively).

Conclusion: For FDG-PET images on PET/MR, attenuation correction using ZTE-MRI had superior accuracy to an atlas-based method in classification for dementia. ZTE maintains the diagnostic accuracy for AD.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7725704PMC
http://dx.doi.org/10.3389/fnins.2020.569706DOI Listing

Publication Analysis

Top Keywords

zte atlas
24
pet score
16
diagnostic accuracy
12
adni dataset
12
mrac methods
12
atlas group
12
pet
9
maintains diagnostic
8
alzheimer's disease
8
attenuation correction
8

Similar Publications

Tumor-to-normal ratio (T/N) measurement of F-FBPA is crucial for patient eligibility to receive boron neutron capture therapy. This study aims to compare the difference in standard uptake value ratios on brain tumors and normal brains using PET/MR ZTE and atlas-based attenuation correction with the current standard PET/CT attenuation correction. Regarding the normal brain uptake, the difference was not significant between PET/CT and PET/MR attenuation correction methods.

View Article and Find Full Text PDF

Composite attenuation correction method using a Ge-transmission multi-atlas for quantitative brain PET/MR.

Phys Med

May 2022

Radiology & Molecular Imaging, Department of Surgical Sciences, Uppsala University, Uppsala, Sweden; Medical Physics, Uppsala University Hospital, Uppsala, Sweden.

In positron emission tomography (PET), Ge-transmission scanning is considered the gold standard in attenuation correction (AC) though not available in current dual imaging systems. In this experimental study we evaluated a novel AC method for PET/magnetic resonance (MR) imaging which is essentially based on a composite database of multiple Ge-transmission maps and T1-weighted (T1w) MR image-pairs (composite transmission, CTR-AC). This proof-of-concept study used retrospectively a database with 125 pairs of co-registered Ge-AC maps and T1w MR images from anatomical normal subjects and a validation dataset comprising dynamic [C]PE2I PET data from nine patients with Parkinsonism.

View Article and Find Full Text PDF

Background: Attenuation correction (AC) is an important topic in PET/MRI and particularly challenging after brain tumor surgery, near metal implants, adjacent bone and burr holes. In this study, we evaluated the performance of two MR-driven AC methods, zero-echo-time AC (ZTE-AC) and atlas-AC, in comparison to reference standard CT-AC in patients with surgically treated brain tumors at C-methionine PET/MRI.

Methods: This retrospective study investigated seven postoperative patients with neuropathologically confirmed brain tumor at C-methionine PET/MRI.

View Article and Find Full Text PDF

Deep learning (DL) strategies applied to magnetic resonance (MR) images in positron emission tomography (PET)/MR can provide synthetic attenuation correction (AC) maps, and consequently PET images, more accurate than segmentation or atlas-registration strategies. As first objective, we aim to investigate the best MR image to be used and the best point of the AC pipeline to insert the synthetic map in. Sixteen patients underwent a 18F-fluorodeoxyglucose (FDG) PET/computed tomography (CT) and a PET/MR brain study in the same day.

View Article and Find Full Text PDF

Background: A valid photon attenuation correction (AC) method is instrumental for obtaining quantitatively correct PET images. Integrated PET/MR systems provide no direct information on attenuation, and novel methods for MR-based AC (MRAC) are still under investigation. Evaluations of various AC methods have mainly focused on static brain PET acquisitions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!