A Pilot Study of Automated Pupillometry in the Treatment of Opioid Use Disorder.

J Addict Med

Department of Ophthalmology, Boston Medical Center, Boston University School of Medicine, Boston, MA (CEP, MG, NS, SS, NT, MGF), Department of Neurology, Boston Medical Center, Boston University School of Medicine, Boston, MA (CEP), Department of Biostatistics, Boston Medical Center, Boston University School of Medicine, Boston, MA (HC), Department of Medicine, Boston Medical Center, Boston University School of Medicine, Boston, MA (MKP-O, ZMW).

Published: December 2021

Introduction/background: A rough, visual estimate of pupil size is used in grading the severity of opioid withdrawal. Few studies have examined the clinical utility of more precise automated pupillometry measurements.

Methods: This prospective cohort study enrolled 27 patients receiving opioid agonist therapy (OAT) to treat cravings or withdrawal during an acute hospitalization. Six sets of automated pupillometry measurements were obtained at regular intervals before and after administration of OAT. Clinical Opiate Withdrawal Scale measurements were performed pre and post OAT. Primary outcomes included pupil size in dark and bright illumination (mm). Latency of the pupillary light response (s), constriction and dilation velocity (mm/s), and percent constriction (%) were secondary outcomes.

Results: The mean predosing pupil size in dark and bright illumination was 4.33 ± 1.40 mm and 2.96 ± 0.79 mm, respectively. A significantly decreased mean pupil size was first detected at 15 minutes postdosing (4.01 ± 1.34 mm, P = 0.0115 for dark illumination; 2.71 ± 0.72 mm, P = 0.0003 for bright illumination) and this reduction in pupil size persisted at later postdosing timepoints. Those with Clinical Opiate Withdrawal Scale <5 after dosing had a greater decrease in dark pupil size (10.6% ± 13.2 vs 3.2% ± 3.2, P = 0.043). There was no significant change in the remaining pupil reactivity parameters.

Conclusions: Automated pupillometry demonstrated a small but significant change in mean pupil size that occurred within 15 minutes of OAT dosing and was associated with low withdrawal scores. This pilot may inform future work to incorporate pupillometry measurement into OAT dosing assessments.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8562941PMC
http://dx.doi.org/10.1097/ADM.0000000000000794DOI Listing

Publication Analysis

Top Keywords

pupil size
20
automated pupillometry
12
bright illumination
12
clinical opiate
8
opiate withdrawal
8
withdrawal scale
8
size dark
8
dark bright
8
pupil
5
size
5

Similar Publications

Aperiodic Pupil Fluctuations at Rest Predict Orienting of Visual Attention.

Psychophysiology

January 2025

Department of Psychology, Hangzhou Normal University, Hangzhou, Zhejiang, China.

The aperiodic exponent of the power spectrum of signals in several neuroimaging modalities has been found to be related to the excitation/inhibition balance of the neural system. Leveraging the rich temporal dynamics of resting-state pupil fluctuations, the present study investigated the association between the aperiodic exponent of pupil fluctuations and the neural excitation/inhibition balance in attentional processing. In separate phases, we recorded participants' pupil size during resting state and assessed their attentional orienting using the Posner cueing tasks with different cue validities (i.

View Article and Find Full Text PDF

Visual Performance Correlation with Corneal Aberrometric Profile and Pupil Size After Implantation of a Trifocal Hydrophobic IOL.

Clin Ophthalmol

December 2024

Département de Recherche Clinique, Institut Ophtalmologique de l'Ouest (IOO) Jules Verne, Nantes, France.

Purpose: To evaluate clinical outcomes obtained after cataract surgery involving the implantation of a trifocal hydrophobic intraocular lens (IOL) and to determine if pupil size and the corneal aberrometric profile correlate to visual acuity at different distances.

Methods: 49 patients (98 eyes) underwent bilateral cataract surgery with the placement of FineVision HP IOLs for presbyopia and were assessed at 1- and 3- to 6-months post-surgery. Postoperatively, refraction, monocular and binocular uncorrected and corrected distance visual acuity (UDVA and CDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA) and the binocular defocus curve were measured.

View Article and Find Full Text PDF

Photic phenomena are more pronounced in presbyopia-corrected than in monofocal intraocular lens (IOL), causing dissatisfaction after cataract surgery. Photic Phenomena Test (PPT) quantifies photic phenomena in eyes with two types of presbyopia-corrected IOL. We examined the relationship between preoperative eye shape and pupil diameter.

View Article and Find Full Text PDF

Schwannomas (SCs) are benign tumors composed of neoplastic Schwann cells and are relatively uncommon intracranially. Although these tumors are frequently associated with neurofibromatosis type 2 (NF2), they may also arise idiopathically, and their pathogenesis remains poorly understood. A 70-year-old Caucasian man presented with a two-month history of vertigo, gait imbalance, and decreased visual acuity in the left eye accompanied by photophobia, nausea, vomiting, and occasional headaches.

View Article and Find Full Text PDF

Microtextured microneedles are tiny needle-like structures with micron-scale microtextures, and the drugs stored in the microtextures can be released after entering the skin to achieve the effect of precise drug delivery. In this study, the skin substitution model of Ogden's hyperelastic model and the microneedle array and microtexture models with different geometrical parameters were selected to simulate and analyse the flow of the microtexture microneedle arrays penetrating the skin by the finite-element method, and the length of the microneedles was determined to be 200 μm, the width 160 μm, and the value of the gaps was determined to be 420 μm. A four-pronged cone was chosen as the shape of microneedles, and a rectangle was chosen as the shape of the drug-carrying microneedle.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!