Background: Modularity of metal-on-metal (MoM) implants has come under scrutiny due to concerns regarding additional sources of metal debris. This study is a retrieval analysis of implants from the same manufacturer with the same MoM bearing surface. The difference between the implants was presence or absence of modular junctions.
Methods: This is a retrospective study of 31 retrieved implants from 31 patients who received a Conserve Wright Medical MoM hip prosthesis. The 31 implants consisted of 16 resurfacings and 15 implants with modular junctions; 4 conventional THAs and 11 modular-neck THAs.
Results: 43% of pre-revision MRI scans performed on resurfacing implants and 91% performed on the modular implants illustrated evidence of an adverse local tissue reaction. There was no difference in pre-revision blood metal ion levels or bearing surface wear between the resurfacings and modular implants. The neck-head tapers of the modular group showed low levels of material loss. However, the neck-stem tapers showed increased severity of corrosion and material loss.
Conclusions: The modular implants had an increased incidence of adverse local tissue reaction. This could be related to the presence of modular junctions, particular the neck-stem junction which showed increased susceptibly to corrosion.