Response of the ArcCHECK® device at 6 MV and 15 MV for VMAT and IMRT quality control.

Phys Med

Instituto de Investigación Biosanitaria (ibs.GRANADA), Complejo Hospitalario Universitario de Granada/Universidad de Granada, E-18016 Granada, Spain; Departamento de Física Atómica, Molecular y Nuclear, Universidad de Granada, E-18071 Granada, Spain.

Published: December 2020

Purpose: To study the response of the ArcCHECK® device as VMAT and IMRT verification system.

Methods: Various tests analyzing the linearity, the repeatability and the angular dependence of the device response, its dependence with the pulse repetition rate and the leakage losses were performed. The long-term response in dose measurements and the uniformity of the detectors conforming the system were controlled using a statistical process control program. The Elekta Infinity™ 6 and 15MV photon beams were used.

Results: The device showed excellent repeatability and linearity. The differences between the responses obtained for any pair of angular incidences were less than 2%. The absorbed dose increased by 3% when the pulse repetition rate varied from 50 to 600MU/min. Results are in overall agreement with those found in previous works for the ArcCHECK®, in which a reduced number of the device diodes were analyzed, and for the MapCheck®, an older 2D device that used the same diodes. Charge losses were found to be negligible except for some of the diodes of the device. The statistical process control program is a very useful tool to control the correct functioning of the device in the long term.

Conclusions: The results of the analysis carried out indicate that the working and stability conditions of the ArcCHECK® device are adequate for its purpose. The dependence with the pulse repetition rate should be considered when VMAT or similar treatments are evaluated. A control program for the statistical monitoring of the device would be desirable and useful.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2020.11.020DOI Listing

Publication Analysis

Top Keywords

arccheck® device
12
pulse repetition
12
repetition rate
12
control program
12
device
10
response arccheck®
8
vmat imrt
8
dependence pulse
8
statistical process
8
process control
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!