Introduction: Whether sirolimus-eluting stents constituted with ultrathin-strut and biodegradable polymers (BP-SESs) can achieve a preferable effect over current drug-eluting stents with durable polymers (DP-DESs) remains highly controversial. The aim of this analysis based on randomized controlled trials (RCTs) was to detect the clinical and angiographic differences between ultrathin (defined as a strut thickness <70 µm) BP-SESs and DP-DESs.
Evidence Acquisition: We searched seven databases to identify eligible articles. Late lumen loss (LLL) and target lesion failure (TLF) were assessed as the primary endpoints for angiographic and clinical outcomes, respectively.
Evidence Synthesis: Nineteen articles containing thirteen RCTs with 14801 patients were analyzed. For the 9-month angiographic outcomes, similar results were discovered between BP-SESs and DP-DESs in terms of in-stent LLL (mean difference [MD]: -0.02 [-0.05, 0.01], P=0.23), in-segment LLL (MD: -0.01 [-0.04, 0.03], P=0.74), in-stent minimum lumen diameter (MLD) (MD: -0.01 [-0.06, 0.04], P=0.72), in-segment MLD (MD: -0.01 [-0.06, 0.05], P=0.75), in-stent diameter stenosis (DS) (MD: -1.10 [-3.36, 1.15], P=0.34), in-segment DS (MD: -0.78 [-1.97, 0.40], P=0.20), in-stent binary restenosis (BR) (risk ratio [RR]: 2.27 [0.99, 5.21], P=0.05) and in-segment BR (RR: 1.46 [0.78, 2.75], P=0.24). Regarding the 12-month clinical outcomes, there was a significant decrease in TLF and a trend of a lower incidence of target vessel failure (RR: 0.89 [0.78,1.01], P=0.08), myocardial infarction (MI) and target vessel MI.
Conclusions: With similar angiographic results, BP-SESs appeared to be superior to DP-DESs with better clinical prognoses, especially for female patients, patients with STEMI and ACS and patients without diabetes. More high-quality randomized controlled trials are needed to confirm these results.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.23736/S0021-9509.20.11620-3 | DOI Listing |
Catheter Cardiovasc Interv
January 2025
State Key Laboratory of Frigid Zone Cardiovascular Diseases, Cardiovascular Research Institute and Department of Cardiology, General Hospital of Northern Theater Command, Shenyang, China.
Background: Biodegradable polymer stents may reduce the risk of neoatherosclerosis and stent thrombosis. Limited data is available for biodegradable polymer sirolimus-eluting stent (BP-SES) and durable polymer drug-eluting stents (DP-EES) in chronic total occlusions (CTO).
Aim: This study was to evaluate healing patterns of BP-SES versus DP-EES in CTO at 3 and 13 months based on optical coherence tomography (OCT).
Am Heart J
January 2025
Department of Cardiology, Odense University Hospital, Odense, Denmark; University of Southern Denmark, Odense, Denmark.
Rationale: The biodegradable polymer BioMatrix Alpha™ stent contains biolimus A9 drug which is sirolimus derivative increase in lipophicity. The biodegradable polymer sirolimus eluting Combo™ stent is a dual-therapy sirolimus-eluting and CD34+ antobody coated stent capturing endothelial progenitor cells (EPCs).
Hypothesis: The main hypothesis of the SORT OUT XI trial was that the biodegradable polymer biolimus A9 BioMatrix Alpha ™ stent is noninferior to the biodegradable polymer sirolimus eluting Combo™ stent in an all-comers population with coronary artery disease undergoing percutaneous coronary intervention (PCI).
Am J Cardiol
January 2025
Department of Cardiovascular Medicine, Osaka University Graduate School of Medicine, Suita, Japan.
A dual-therapy sirolimus-eluting and CD34+ antibody-coated Combo Stent (DTS) has been developed to enhance endothelization and capture endothelial progenitor cells; however, vessel responses following DTS implantation remain unclear. Therefore, we evaluated early- and mid-term intravascular characteristics of DTS using intravascular imaging modalities. This multicenter, prospective, observational study enrolled 88 patients (95 lesions) who underwent DTS (43 patients, 48 lesions) or sirolimus-eluting Orsiro stent (SES, 45 patients, 47 lesions) implantation.
View Article and Find Full Text PDFFront Med (Lausanne)
January 2025
Department of Nephrology and Dialysis, Brugmann University Hospital, Université Libre de Bruxelles (ULB), Brussels, Belgium.
Background: The arteriovenous fistula (AVF) is the preferred vascular access for hemodialysis. AVF stenosis is a common complication, often requiring balloon angioplasty. For recurrent stenosis, AVF stenting may be an option.
View Article and Find Full Text PDFJACC Cardiovasc Interv
January 2025
Department of Cardiology, Fu Wai Hospital, National Center for Cardiovascular Diseases of China, Beijing, China. Electronic address:
Background: First-generation bioresorbable scaffolds (BRS) increased risks of stent thrombosis and adverse events. The Bioheart scaffold is a new poly-L-lactic acid-based BRS.
Objectives: This study sought to evaluate the efficacy and safety of the BRS in patients with coronary artery disease.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!