Background and purpose - Modular-neck hip stems have been identified with corrosion-related problems from the neck-stem junction. We report an ongoing varus deformity of modular-neck hip stems with simultaneous metal ion release observed during a study comparing the migration of modular vs. standard hip stems.Patients and methods - We followed 50 patients with modular and 25 with standard neck stems using radiostereometry (RSA). At 5-year follow-up, we noted a compromised integrity of the modular stem with varus deformity in the neck-stem interface. Changes in head-tip distance as well as whole-blood ion concentration and MRI findings were analyzed. The modular stems were followed further up to 8 years.Results - The head-tip distance decreased continuously by 0.15 mm per year resulting in 1.2 (95% CI 1.0-1.4) mm at 8 years for modular stems, while for the standard stems at 5 years, the decrease was 0.09 (CI 0.0-0.2) mm or 0.02 mm/year. For the modular stems, the reduction in head-tip distance correlated to the increase in whole-blood cobalt concentration at 8 years but not to the MRI grading of tissue reactions. At 5 years, cobalt levels were 4.9 µg/L for modular stems and at 8 years 4.8 µg/L, whereas for standard stems this was 1.0 µg/L. After 8 years, 9 of 72 stems had been revised for different reasons, but only 1 with obvious adverse local tissue reaction (ALTR).Interpretation - We present a surprisingly large progressive deformation at the modular neck-stem junction, but so far without a definite clinical problem. Even the femoral head seems to show slight compression onto the taper over time. A high rate of revisions for the modular type of this stem has raised general concerns, and it has been recalled from the market.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7919878PMC
http://dx.doi.org/10.1080/17453674.2020.1853387DOI Listing

Publication Analysis

Top Keywords

modular stems
16
varus deformity
12
modular-neck hip
12
hip stems
12
head-tip distance
12
stems
11
modular
9
metal ion
8
ion release
8
mri findings
8

Similar Publications

Total hip arthroplasty (THA) is a very successful operation. Once the problem of implant fixation was solved with the use of bone cement, the next development steps focused on improving the bearing. Weber, a Swiss surgeon, introduced the first modular heads in THA.

View Article and Find Full Text PDF

Background: Anatomic total shoulder arthroplasty are highly successful procedures for treatment of glenohumeral arthritis to reduce pain, improve range of motion, and overall quality of life. However, the long-term survivorship of the implant systems is less widely documented in the existing literature. The purpose of this study was to establish the long-term patient outcomes and identify factors influencing the postoperative implant survivorship of total anatomic shoulder arthroplasty devices/procedures utilizing the short-stemmed prosthesis and hybrid glenoid components of the Comprehensive Shoulder System (Zimmer Biomet, Warsaw, IN, USA) at an average of ten years.

View Article and Find Full Text PDF

Finding effective antibiotics against multi-resistant strains of bacteria has been a challenging race. Linker-Evolved-Group-Optimized-Lipophosphonoxins (LEGO-LPPOs) are small modular synthetic antibacterial compounds targeting the cytoplasmic membrane. Here we focused on understanding the reasons for the variable efficacy of selected LEGO-LPPOs (LEGO-1, LEGO-2, LEGO-3, and LEGO-4) differing in hydrophobic and linker module structure and length.

View Article and Find Full Text PDF

Background: Periprosthetic femoral fractures (PFF) are a challenging complication of hip arthroplasty surgery, posing a high risk of morbidity, mortality and reoperation. The Vancouver Classification describes a B2 PFF around a loose stem with sufficient bone stock. In recent years, the number of B2 PFFs and cementation of femoral stems have increased substantially.

View Article and Find Full Text PDF

Use of semi-constrained total knee arthroplasty in gonarthrosis with collateral ligament insufficiency: Clinical and functional outcomes.

Rev Esp Cir Ortop Traumatol

December 2024

Departamento de Traumatología y Cirugía Ortopédica, Hospital Clínico San Carlos, Madrid, España; Departamento de Cirugía, Universidad Complutense, Madrid, España.

Introduction: Achieving stability in total knee arthroplasty (TKA) is crucial for long-term implant survival. In cases of severe deformity or ligament laxity, constrained implants may be required. Traditionally, increasing constraint involved intramedullary stems.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!