A simple method for measuring electron drift velocity in gases using grid ionization chamber.

Appl Radiat Isot

Department of Nuclear Science and Technology, School of Energy and Power Engineering, Xi'an Jiaotong University, Xi'an, 710049, China.

Published: February 2021

A simple method for measuring the electron drift velocity in gases with a given electric field using a grid ionization chamber is proposed and demonstrated. By collimating incident α particles that are perpendicular to the electric field, the drift velocity can be derived easily using the electron drift distance from primary ionization to a grid divided by the time interval between the cathode and anode signal starting times. These experimental settings can avoid additional signal processing of signals and reduce the effect of electron diffusion. Using this method, the measurement of electron drift velocities in 90% Ar + 10% CO is presented. Measured results agree well with the simulated values and with existing experimental results.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2020.109496DOI Listing

Publication Analysis

Top Keywords

electron drift
16
drift velocity
12
simple method
8
method measuring
8
measuring electron
8
velocity gases
8
grid ionization
8
ionization chamber
8
electric field
8
electron
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!